SDM – Slurry Density Meter
44
www.rhosonics.com
CONFIGURATION
3 Load liquid
The liquid calibration is loaded from a liquid file this file starts with RLS and has the csv file-
extension. This file needs to be on this exact location:
C:\Users\”USERNAME”\Documents\LabVIEW Data\
When this location doesn’t exist please create this location.
Press Load liquid
When Lamp Liquid loaded is green the settings are loaded
4 Read liquid
To read the liquid from the analyzer a liquid file is needed. The name of this file starts with RLS and has
the csv file-extension.
This file needs to be on this exact location:
C:\Users\”USERNAME”\Documents\LabVIEW Data\
The file is needed to provide the description.
When this location doesn’t exist please create this location.
Press Read liquid
When the descriptions and values in the analyzer are visible and the file column is grey, the liquid
calibration is read from the analyzer
Configuring minimum and maximum values for the 4-20 mA output
Open “SETTING.csv” file in Excel
Change the values for “PV lower range value” and “PV upper range value”
Save the file (keep .csv extension)
Press Load settings
When Lamp Settings loaded is green the settings are loaded
The minimum and maximum values for the 4-20 mA output are now configured.
Changing Primary, Secondary, Ternary and Quaternary values for HART output
Open “SETTING.csv” file.
Change the values for “primary/secondary/ternary/quartinary dynamicVarConfig”, all available
quantities have been included in the csv-file
Save the file (keep .csv extension)
Press Load settings
When Lamp Settings loaded is green the settings are loaded
The Primary, Secondary, Ternary and Quaternary values for HART output are now set.