Calibration
It is necessary to inspect the condition of the calibration tile and instrument optics
before each calibration.
Inspect the tile and instrument optics before calibration.
Inspecting and cleaning the instrument optics
Inspecting and cleaning the instrument optics
Inspecting and cleaning the instrument optics
Inspecting and cleaning the instrument optics
Any dust or debris on the optic should be blown from the lenses using dry clean
air, the optics must not be touched. If there are any permanent marks or scratches
on the lenses, the instrument is no longer suitable for measuring and should be
returned to an authorised Rhopoint service center.
Inspecting and cleaning the calibration tile
The calibration tile must be perfectly clean from smears and scratches before
attempting calibration. Fingerprints and dust can be removed with the supplied
optic cleaning cloth.
Scratched or damaged tiles
Scratched or damaged tiles are not suitable for calibration. The instrument and tile
must be returned to an approved Rhopoint service center for tile replacement and
recalibration.