Instrument Measurement Verification
The Novo-Haze TX is calibrated at the factory using traceable calibration artefacts. If
the measured value of the checking standard does not match the assigned values
(within the tolerances), please follow the following steps:
1. Ensure the checking standard is in date.
2. Ensure the standard being used is not damaged. Any dust, scratches or other
marks will alter the reading taken by the instrument.
3. Ensure the light baffle is down when taking a reading.
4. Ensure the instrument is tared correctly. This can be checked by taking a
reading without any sample after performing a tare - the haze value should be
0% and the transmittance 100%.
Sample Holders
Custom sample holders and adaptors, to ensure correct product placement, are
available on request. Please contact Rhopoint Instruments with your requirements.