RHG Products Company 

RO Only Option 

RO filters greatly reduce water TDS without the aid of DI filters.  

Under pump pressure this reduction will be greater than under tap 

pressure. Windows can be washed spot free with water up to 10ppm 

TDS. Non-Glass surfaces can typically be washed with water up to 

40ppm TDS.  

 

1. To run your system in RO Only Mode open the 1/2” grey valve labeled 

‘R.O.’ and close the 1/2” grey valve labeled ‘D.I.’ 

 

2. If the system is not producing water of acceptable purity in RO Only 
Mode for your application switch to DI mode by closing the 1/2” grey 

valve labeled ‘R.O.’ and opening the 1/2” grey valve labeled ‘D.I.’  

 

Water will flow from different outlets depending on which valve you 

have open.  

ALWAYS BE SURE ONE GREY 

VALVE IS OPEN DURING PRODUCTION 

 

TDS CREEP 

 

** PREVENT TDS CREEP AND EXTEND DI LIFE ** 

For a few moments after RO membranes begin production, they fail to 

reject the percentage of dissolved solids they are designed to. Each time 

you start your system, run it in RO Only mode until your RO membranes 

“warm up” and are reducing product water TDS by about 90%. If you 

now need to clean with pure water, switch to D.I. mode.  

 

 

 

 

 

 

Содержание H2PRO

Страница 1: ...RHG Products Company RHG 3 4 Stage Water Powered H2PRO User Manual ...

Страница 2: ...RHG Products Company ...

Страница 3: ...ect for any damage that may have occurred during shipping 2 Be sure to inventory all items that were to be included with your order 3 Any damage or missing components must be reported within 24 hours to the supplier of your system 4 Your H2PRO is shipped complete in 1 box ...

Страница 4: ...nitial set up and when you change your carbon filter 6 As water begins to flow out of the bypass valve blue valve below the inlets and outlet slowly close the bypass valve all the way Although the valve is fully closed water will still flow out of it This is normal 7 It will take a few minutes for all of the air to purge out of your system before water makes its way out of the reel 8 After the ini...

Страница 5: ... acceptable purity in RO Only Mode for your application switch to DI mode by closing the 1 2 grey valve labeled R O and opening the 1 2 grey valve labeled D I Water will flow from different outlets depending on which valve you have open ALWAYS BE SURE ONE GREY VALVE IS OPEN DURING PRODUCTION TDS CREEP PREVENT TDS CREEP AND EXTEND DI LIFE For a few moments after RO membranes begin production they f...

Страница 6: ...longer than 2 weeks close your DI valve and open your RO only valve Run the RO in flush mode for 4 minutes and RO production mode for 10 minutes before reconnecting the DI cartridge check RO efficiency at this time 3 Turn off the water supply to the unit 4 Disconnect the water supply It is important to always flush the system after each use Failure to do so will decrease the filters life ...

Страница 7: ...carbon sediment cartridge needs to be replaced every 6 months to protect the RO membranes Failure to do so will void your warranty The job of the carbon filter is to remove the chlorine It is possible to check to see if your carbon filter is still working efficiently with a chlorine test kit Your H2PRO system is a water purification system and as such it needs to be maintained Running water throug...

Страница 8: ...s installed There is an o ring on the filter This o ring should go at the top end of the housing Insert the filter from the opposite end first Be sure to run water through the system without any booster motor pump assembly powered on to purge all air out of the system before or you risk damaging the filters Reassemble filter housing and reconnect hoses You may need to relubricate your filter housi...

Страница 9: ...RHG Products Company ...

Страница 10: ...RHG Products Company ...

Страница 11: ...er coming from the water supply If your pressure is below 50 PSI you will want to consider a booster pump to increase the pressure coming into the system If you are not getting adequate flow at the brush of your waterfed pole check your pressure gauge to see that your incoming PSI is 50 or higher If it is you may have a blockage elsewhere in your set up Any alterations to your system will void its...