background image

Qstick User Manual 

 

16

For your Qstick spectrometer, this dark spectrum has already been taken, is stored in the 
device and automatically subtracted from the measured spectrum each time you take a 
new spectrum. Because the dark spectrum depends on the exposure time in a non-linear 
way, there are actually multiple dark spectra stored in the device for a certain number of 
exposure  times.  The  dark  spectrum  that  is  needed  for  a  chosen  exposure  time  is  then  
obtained by interpolating between these dark spectra. 

In  case  you  find  that  the  spectra  you  get  are  not  zero  on  the  average,  even  if  no  light 
enters the spectrometer, then you probably need to recalibrate the dark spectra. You can 
also  check  the  dark  spectrum  calibration  using  the  ‘‘Calibrate  Dark  Spectrum’’  window 
(see below). 

To  recalibrate  the  dark  spectra, simply  close  the  spectrometer entrance,  make sure that 
no  light  enters  the  device  and  choose  ‘‘Calibrate  Dark  Spectrum’’  from  the  Tools  menu. 
Then  click  the  ‘‘Take’’  button  and  go  for  a  coffee.  After  a  few  minutes  the  software  has 
taken dark spectra for a series of exposure times. You can check the new dark spectra by 
clicking on the ‘‘Test Button’’.  While the test is running, you can in the lower diagram the 
spectra  taken  without  (red)  and  with  (green)  dark  spectrum  correction.  In  the  upper 
diagram  you  can  see  the noise  level,  also  without  (red)  and  with  (green)  dark  spectrum 
correction for a series of exposure times. You can see that the noise level of the raw data 
(red) increases significantly with increasing exposure time. With dark spectrum correction 
(green)  however,  the  noise  level  should  remain  pretty  much  at  the  same  level  for 
exposure times less than 1 second and increase slower for exposure times larger than 1 
second. If you find this to be the case, close the window by clicking on the ‘‘OK’’ button 
and  the  new  dark  spectrum  calibration  is  saved  into  the  device.    Otherwise  you  may 
calibrate again or click ‘‘Cancel’’. 

Spectral Sensitivity 

Like in all spectrometers, the sensitivity of the Qstick spectrometer is not uniform across 
the  spectrum,  but  depends  on  the  wavelength.  This  is  called  ‘‘Spectral  Sensitivity’’  and 
can be corrected by dividing each value of the spectrum by the corresponding sensitivity. 
This  sensitivity  calibration  is  also  stored  in  the  device.  However,  it  is  not  always  used 
when  taking  spectra,  because  if  the  sensitivity  is  corrected,  the  noise  level  is  unevenly 
distributed  across  the  spectrum.  Instead,  you  can  choose  to  use  the  spectral  sensitivity 
correction by selecting the ‘‘lightmeter’’ button in the Exposure Toolbar. 

If  you  would  like  to  recalibrate  the  spectral  sensitivity,  you  need  a  light  source  with  a 
continuous  spectrum  (like  a  tungsten  halogen  lamp)  and  a  calibrated  spectrum  for  this 
lamp. 

To  recalibrate,  take  a  spectrum  of  your  light  source  with  spectral  sensitivity  correction 
turned off. Make sure the spectrum contains as little noise as possible by using a strong 
light  source,  a  suitable  exposure  time  and  use  Temporal  Averaging.  Then  choose 
‘‘Calibrate  Spectral  Sensitivity’’  from  the  Tools  menu.  Click  on  the  ‘‘Load  reference  data’’ 
button  to  load  the  calibrated  reference  spectrum  (in  other  words:  a  spectrum  file  that 
shows how your light source really looks like). Based on the measured spectrum and this 

Содержание q stick

Страница 1: ...User Manual ...

Страница 2: ...eme GmbH Product specifications and descriptions in this manual are subject to change without notice RGB Lasersysteme GmbH will not be responsible for errors and omissions in this manual or for direct or incidental damages in connection with the use of this device or information 2012 RGB Lasersysteme GmbH Donaupark 13 93309 Kelheim Germany Telephone 49 9441 1750 33 0 Website http www rgb laser com...

Страница 3: ...m Display 8 3 5 Working with spectra 9 3 6 File Operations 10 3 7 Spectrum Analysis 11 3 8 Absorption Reflection and Transmission Measurements 12 3 9 Calibration 14 4 Software Development Kit 18 4 1 Prerequisites 18 4 2 Getting started 18 4 3 Classes Overview 18 4 4 Adding the SDK library to your project 19 4 5 Taking a spectrum 20 4 6 Deploying your project 20 5 Technical Support 21 5 1 Getting H...

Страница 4: ......

Страница 5: ...tick Hardware Features o Subminiature design o Spectral resolution of 1 2 nm o 2500 pixel CCD sensor o Optical input via SMA fiber o Controlled and powered via USB 2 0 Waves Software Features o Take and display series of spectra o Dynamic exposure control with dark spectrum interpolation o Import most ASCII based file formats o Export as ASCII table to almost any numerical analysis software o Comp...

Страница 6: ...equirements o Windows 8 7 Vista or XP 32 or 64 bit o NET Framework Version 3 5 available as a free download o USB 2 0 1 3 Package Contents o Qstick spectrometer o Waves spectroscopy software device driver and this manual as PDF on the USB memory drive included in the device o Printed user manual Visit us at www rgb laser com if you need any accessories for your spectrometer ...

Страница 7: ...ebsite for software updates The device driver for the Qstick spectrometer is not fully certified yet If Windows displays a warning please choose to install the driver anyway You can be assured that is does not contain any malicious software The USB port needs to supply an electrical current of 200 mA so you need to either connect the spectrometer directly to your computer or to an externally power...

Страница 8: ...tivity correction and the optical power was made with a 200 µm fiber optical cable If you use a different fiber or especially if you just point the spectrometer towards the light source the measured spectrum might not be accurate If you have a calibrated light source you can always recalibrate the spectral sensitivity for any optical setup see chapter 3 9 If you focus the light directly onto the e...

Страница 9: ...trometer should show up as Qstick in the Measurement and Control branch If your spectrometer is marked with an exclamation or question mark please remove the device uninstall the software from the Windows Control Panel restart the computer and install the software again If your spectrometer is not listed at all in the Device Manager please try a different USB port If the device does not work on on...

Страница 10: ... measurements absorption irradiance scope and so on through a consistent and clear interface o Waves favors a simple self explanatory terminology instead of complicated terms that need to be explained For example we use Exposure time instead of Integration time because the average user doesn t really know what s being integrated here o There are no wizards to help you with complicated tasks becaus...

Страница 11: ...tinuous measurement You can increase the exposure time in order to decrease the noise level If the amount of light detected during the exposure is too large for the CCD detector inside the spectrometer the spectrum appears saturated and a red OVERLOAD warning appears at the top of the diagram In this case you need to decrease the exposure time You can further increase the sensitivity by averaging ...

Страница 12: ...y The Main Diagram Panel displays all spectra that are marked by a check mark in the Spectrum List If no spectrum is checked in the list then the currently selected spectrum marked by a gray bar is displayed You can check and uncheck all spectra at once by clicking on the corresponding button or on top of the Spectrum List The peaks in the spectra are automatically determined This not only works p...

Страница 13: ...eep this spectrum and take another one just click on the single push pin button This prevents the current spectrum from being overwritten A new spectrum you take is then appended to the list Giving the current spectrum a new name also prevents it from being overwritten You can also choose to keep all new spectra by selecting the button with multiple push pins in the Exposure Toolbar When you have ...

Страница 14: ...atically determine the file structure and is able to open many different file formats 1 You can print spectra by choosing Print from the file menu If you choose the page orientation as Landscape the spectrum diagram is printed across the whole page If you choose Portrait the diagram is printed on the upper half of the page and the lower half shows selected metadata and values The printing options ...

Страница 15: ...point by the average of the surrounding data points You can choose how many data points to include in the averaging This is also known as Boxcar Averaging Arithmetic Operations This creates a new spectrum by adding subtracting multiplying or dividing two spectra Adjust Wavelength This allows to assigns a new wavelength scale to all spectra in the list based on the parameters for the fourth order p...

Страница 16: ...perature value are used for the corresponding axis In the Strip Chart diagram the currently selected spectrum is marked by a slightly larger point Points corresponding to spectra that are currently not visible in the main window are drawn in a lighter red color Show Loupe This opens a window with a magnified display of the spectrum around the mouse pointer It may be used in order to study small fe...

Страница 17: ...ple for measuring absorption or transmission To take a reference spectrum for reflection spectroscopy you need a white reference sample to calibrate the setup for 100 reflection However all real world reference samples have a reflection of less than 100 and an uneven spectral distribution Therefore for accurate measurements you also need to supply a reference spectrum for your reference sample To ...

Страница 18: ...ed if the temperature has changed considerably After you have recalibrated the spectrometer you can always restore the factory calibration in the Device Properties window from the File menu Calibration Accuracy The wavelength calibration is very accurate and changes only little with time or temperature Therefore most users won t ever need to recalibrate the wavelengths The dark spectrum calibratio...

Страница 19: ...e in the list you need to supply your own calibration lines In this case please take a look at the CalibrationLines xxx dat files in the installation folder on your hard drive and create a new similar file containing the spectrum lines of your light source After you have selected your light source click on Start Fit The software tries to match the peaks in the measured spectrum and the given spect...

Страница 20: ... time With dark spectrum correction green however the noise level should remain pretty much at the same level for exposure times less than 1 second and increase slower for exposure times larger than 1 second If you find this to be the case close the window by clicking on the OK button and the new dark spectrum calibration is saved into the device Otherwise you may calibrate again or click Cancel S...

Страница 21: ...he wavelength range of your spectrometer for example by choosing a suitable light source Then choose Calibrate power from the Tools menu and enter the actual power If you click OK the new power calibration is stored in the device and used for scaling the intensity axis This axis then displays the spectral power density in nW nm In addition you don t have to calibrate the power at the spectrometer ...

Страница 22: ...nality of the SDK library 4 2 Getting started In order to get familiar with the SDK quickly we suggest to 1 Read the next chapter 4 3 to understand the structure of the SDK library 2 Read the basic steps required in order to take a spectrum chapter 4 5 3 Study the sample projects included in the SDK 4 Add the SDK library to your own project chapter 4 4 5 Implement the basic functionality of taking...

Страница 23: ...Studio If you are using one of Microsoft s languages please add both libraries in Visual Studio to your project references in the Solution Explorer window Then import the RgbDriverKit library to your source code files for example in C or C using RgbDriverKit For enhanced IntelliSense support in Visual Studio you may also want to copy the RgbDriverKit xml file into the same folder as the library La...

Страница 24: ...he Status property is SpectrometerStatus TakingSpectrum 6 Read the spectrum with the GetData method The full documentation for these methods and properties can be found in the RgbDriverKit documentation chm help file All methods and some properties may throw an exception as described in the SDK documentation help file Always make sure that these exceptions are properly handled by placing these met...

Страница 25: ...al Support Team via e mail or telephone 5 2 Software Updates Software Updates are expected to be released about twice a year To check for updates you can choose Check for Updates from the Help menu or go to the Support section on our website www rgb laser com These are some of the features planned for future software releases subject to change o More colorimetry values and options o More options f...

Страница 26: ...ector 2500 pixel linear CCD detector A D converter 16 bit 15 MHz Calibration Wavelength sensitivity and multiple dark spectra stored within the device Transfer speed to PC 20 ms per spectrum Digital Interface USB 2 0 Power supply 5 V DC via USB no power adapter required Power consumption typical 200 mA PC operating system Windows 7 Vista XP Mechanical Dimensions 75 8 22 0 10 2 mm Weight 25 g Opera...

Страница 27: ...mbH Donaupark 13 93309 Kelheim declares that the following product Qstick Version 1 0 contains no toxic substances which are specified in the RoHS Directive 2002 95 EG The conformity to RoHS Directive 2002 95 EG is confirmed This declaration applies to all models of the above named product version Kelheim 09 06 2012 Mathias Reichl ...

Отзывы: