4.2. SERIAL COMMUNICATION
Table 4.7: Status and device information commands
Query
Parameter value
Return code
Returnvalue
S?
-
COMMAND_SUCCESS
Current device status (see table below)
R?
-
COMMAND_SUCCESS
Total operating hours [hhhh:mm]
DM?
-
COMMAND_SUCCESS
Manufacturer
DT?
-
COMMAND_SUCCESS
Device name
DS?
-
COMMAND_SUCCESS
Serial number
DO?
-
COMMAND_SUCCESS
Software version
DW?
-
COMMAND_SUCCESS
Emission wavelength in nm
DF?
-
COMMAND_SUCCESS
Available features:
T = Temperature adjustable
H = High speed
integrated
D = Drive mode changeable (
C = Temperature is controlled
M = External analog and digital modulation
A = Modulation inputs can be combined
R = Round beam
L = Wavelock
E = ECDL laser
W = Water-cooled
B = Multi-mode beam
S = Customized
DC?
-
COMMAND_SUCCESS
ACC =
APC =
The result of the status request
S?
is a bit-wise combination of the values listed in
Table 4.8: status request commands
Code
Mnemonic
Description
0x01
LASER_ON
Laser system in active, radiation can be emitted
0x02
(reserved)
0x04
INTERLOCK_OPEN
The interlock is open
0x08
ERROR
Error is active (error can be read with E? query)
0x10
TEMPERATURE_OK
Temperature of laser head is ok
Error status of the laser system
The query
E?
reads the actual active errors of the laser system.
Table 4.9: Error status of the laser system
Query
Parameter value
Return code
Return value
E?
-
COMMAND_SUCCESS
Error code (see table below)
Possible error codes are shown in
17