CHAPTER
4 Maintenance
Page 18 of 23
CA630 In-Service Manual
0510-1116-C
3.
If the control unit indicates an
Assistance Required
alarm, but the button on
the nurse call cord set is in the reset position, check the connection between
the nurse call cord set and the adaptor cable. The ¼” plug on the end of the
nurse call cord set must be fully engaged with the mating jack of the
adaptor cable (see figures below).
Good Connection
Bad Connection
Testing Your System
Test your CA630 System on a monthly basis to verify proper operation.
To test your Fall Management Functionality
1.
Follow the set up instruction in the section “Setting Up The System.”
2.
With the system monitoring weight on the pad, leave the bed or chair.
3.
If the system is operating properly, an alert will sound at the control unit
and the Client computer.
4.
Repeat at several locations on pad.
To test your Nurse Call Functionality
1.
Follow the set up instruction in the section “Setting Up The System.”
2.
With the system set up for Nurse Call, push the button on the Nurse Call
cord set.
3.
If the system is operating properly, an alert will sound at the control unit
and Client computer.