background image

 

  

R E V O L U T I O N A R Y   S C I E N C E  

Incufridge RS-IF-233 

Incufridge Specifications 

 

Model No.: 

 

 

RS-IF-233basic 

Precision: 

 

 

±0.7°C (at 20°C ambient)  

Uniformity: 

 

 

Internal temperature uniformity ±1.0°C (at 20°

 

 

 

 

C ambient)  

Display:   

 

 

3 digits, 0.1°C resolution 

Controller: 

 

 

Digital microcomputer 

Power Requirements: 

 

120-240VAC, 50-60 Hz, 150 W, 12 VDC  

Net Weight: 

 

 

34 lb. 

Outside Dimensions: 

 

15" wide x 23” deep x 15.5" tall 

Inside Dimensions: 

 

12" wide x 16.5" deep x 10.25” tall 

Capacity:  

 

 

33L (1.2 cu. ft.) 

Temperature Range: 

 

Accurately holds from 7°C to 60°C  (at 20°C  

 

 

 

 

ambient)  

 

Warranty: 

 

 

2 year, repair or replacement 

 

Contact Us 

 

Mailing address:   

 

Toll Free: (800) 775-7942 

Revolutionary Science 

 

Fax: (775) 218-7015 

17319 Lake Blvd.   

 

E-mail: 

[email protected]

 

Shafer, MN 55074 USA 

 

Web site: 

www.revsci.com

 

 

 

R E V O L U T I O N A R Y   S C I E N C E  

Incufridge RS-IF-233 

Care and Maintenance 

 

Your Incufridge should be cleaned periodically to ensure optimal performance,  clean-
liness, and to eliminate contamination.  Clean your REVOLUTIONARY        SCIENCE 
RS-IF-233 only when it is not plugged in.  Using a soft cloth soaked in water and 

bleach or mild detergent, wipe the inside and outside of the unit.  Then let air dry or 
wipe with a dry cloth. Be careful not to scratch surfaces with abrasive cleaners or 
dirty cloths. 

 
User Calibration Instructions 

 

Note: To set temperature, follow instructions on page 4. 
 

Setup: 
Set desired temperature on Incufridge and insert calibrated thermocouple or       
thermometer closest to the geometric center of the chamber.  Close door and check 
accuracy of temperature 1 to 2 hours later.  If temperature is off by more than 0.5°C, 
record Incufridge display reading and thermometer reading and calibrate accordingly, 

following these three simple steps. 
 
1. Press up and down buttons simultaneously.  Display should read “CAL.” 

2. After "CAL" times out a new number will display.  This is the calibration off-set 

number.  Press up arrow if you desire to increase the temperature on the display 
reading or down arrow if you wish to decrease the temperature on the display. 

3. Wait for three seconds for the display to return to the temperature reading. 
 
NOTE: Incufridge will maintain adjusted calibration if it is turned off or restarted.  
Calibration may be adjusted anytime. 
 

Temperature Uniformity: 
Geometric extremes (i.e. within an inch or two from the inside of the door, ceiling, 
fan, walls etc) may be less precise. 
For best results, it is recommended to place specimens and samples on the trays, and 

if possible, close to the geometric center. 