Incufridge Specifications
Model No.:
RS-IF-233basic
Precision:
±0.7°C (at 20°C ambient)
Uniformity:
Internal temperature uniformity ±1.0°C (at 20°
C ambient)
Display:
3 digits, 0.1°C resolution
Controller:
Digital microcomputer
Power Requirements:
120-240VAC, 50-60 Hz, 150 W, 12 VDC
Net Weight:
34 lb.
Outside Dimensions:
15" wide x 23” deep x 15.5" tall
Inside Dimensions:
12" wide x 16.5" deep x 10.25” tall
Capacity:
33L (1.2 cu. ft.)
Temperature Range:
Accurately holds from 7°C to 60°C (at 20°C
ambient)
Warranty:
2 year, repair or replacement
Contact Us
Mailing address:
Toll Free: (800) 775-7942
Revolutionary Science
Fax: (775) 218-7015
17319 Lake Blvd.
E-mail:
Shafer, MN 55074 USA
Web site:
www.revsci.com
R E V O L U T I O N A R Y S C I E N C E
Incufridge RS-IF-233
5
Care and Maintenance
Your Incufridge should be cleaned periodically to ensure optimal performance, clean-
liness, and to eliminate contamination. Clean your REVOLUTIONARY SCIENCE
RS-IF-233 only when it is not plugged in. Using a soft cloth soaked in water and
bleach or mild detergent, wipe the inside and outside of the unit. Then let air dry or
wipe with a dry cloth. Be careful not to scratch surfaces with abrasive cleaners or
dirty cloths.
User Calibration Instructions
Note: To set temperature, follow instructions on page 4.
Setup:
Set desired temperature on Incufridge and insert calibrated thermocouple or
thermometer closest to the geometric center of the chamber. Close door and check
accuracy of temperature 1 to 2 hours later. If temperature is off by more than 0.5°C,
record Incufridge display reading and thermometer reading and calibrate accordingly,
following these three simple steps.
1. Press up and down buttons simultaneously. Display should read “CAL.”
2. After "CAL" times out a new number will display. This is the calibration off-set
number. Press up arrow if you desire to increase the temperature on the display
reading or down arrow if you wish to decrease the temperature on the display.
3. Wait for three seconds for the display to return to the temperature reading.
NOTE: Incufridge will maintain adjusted calibration if it is turned off or restarted.
Calibration may be adjusted anytime.
Temperature Uniformity:
Geometric extremes (i.e. within an inch or two from the inside of the door, ceiling,
fan, walls etc) may be less precise.
For best results, it is recommended to place specimens and samples on the trays, and
if possible, close to the geometric center.