Problem 

 Possible Cause 

   Solution

No power or lights 

to Revitive when it is 

switched on.

AC Adapter not switched on 

at the electrical socket or AC 

Adapter not plugged into device 

properly.

Check electrical socket power  

is switched on and the AC Adapter 

is plugged into the device correctly. 

If still not working – contact your 

authorised dealer.

Revitive is not 

vibrating.

REVITIVE IS NOT DESIGNED TO VIBRATE.

The IsoRocker® is not 

"rocking".

It is the muscles in your legs 

that cause the "rocking", the 

IsoRocker® is a pivot.

The IsoRocker® will only rock 

when increased to an intensity 

which causes sufficient calf muscle 

contraction. It may be that you 

cannot comfortably increase the 

intensity high enough until you 

get used to the sensation. It is 

important that intensity is adjusted 

to a level that is comfortable.

The device is too far in front  

of you.

Sit with your knees at a 90 degree 

angle.

When using the 

IsoRocker® on a 

hard floor it makes a 

tapping noise. 

Incorrect positioning of the 

device or too high an intensity  

level. 

Adjust the positioning of the 

device or lower the intensity level 

to reduce the device tapping. 

Alternatively use a floor mat under 

the device to cushion the sound.

My legs are aching  

after treatment.

You may have the       intensity 

on too high a setting and your 

muscles are being overworked. 

Leave adequate time after each 

treatment to allow the muscles 

to recover (just like after vigorous 

exercise!). On your next session 

start on a lower setting (where 

you can feel the mild electrical 

stimulation and it is comfortable) 

and reduce the duration until your 

muscles have acclimatised to the 

stimulation. 

For more information, including Frequently Asked Questions, 

please visit support.revitive.com

Troubleshooting

Содержание ProHealth

Страница 1: ...Device User s Manual CIRCULATION BOOSTER Please read the User s Manual carefully before using this product USE ONLY AS DIRECTED If symptoms persist consult your healthcare professional Model number 5572AQ Neuromuscular Electrical Stimulation Device Single Mode ...

Страница 2: ...setup and using your Revitive system go to support revitive com For best results Use daily 30 60 min Intensity 40 for 60 days 2 year warranty To activate your free 2 year warranty please register your device at support revitive com ...

Страница 3: ...uards 9 16 Who should NOT use Revitive 9 Warnings and Cautions 10 14 Safety Precautions 15 Adverse Reactions 16 Instructions for use 17 22 Step 1 Setting up Revitive for use 17 Step 2 Using the SoftTouch foot pads 18 21 Step 3 Cleaning and storing 22 Troubleshooting 23 24 Technical Specifications 25 28 Warranty 29 Table of contents ...

Страница 4: ...heck that the following components are provided If you think anything is missing please contact us using the helpline numbers on the back of this booklet AC DC power adapter Cable length 180cm Revitive ProHealth Remote control AAA batteries x 2 1 2 3 4 ...

Страница 5: ...5 Control panel display AC DC power adapter socket Parts and controls SoftTouch Pads IsoRocker Power ON OFF Carry Handle ...

Страница 6: ...6 Power ON OFF Remote control Control panel Intensity level Intensity Control Parts and controls Time remaining Intensity Control ...

Страница 7: ...pasm Increase of blood flow circulation Immediate post surgical stimulation of calf muscles to prevent venous thrombosis For users with conditions or diseases that are associated with impaired poor blood flow in the legs ankles feet NMES though the foot pads is intended for use as an adjunctive treatment as an addition to your existing treatment to temporarily reduce lower extremity pain swelling ...

Страница 8: ...s comfortable for you to use It features an IsoRocker that allows for natural ankle movement when applying the EMS to your feet Ankle movement is a key factor in increasing circulation in the lower legs You can choose to enable or disable the IsoRocker at any time Revitive is designed to be easy to use and also comes with a remote control which means you do not have to reach down to adjust setting...

Страница 9: ...ombosis DVT DVT is a blood clot and usually occurs in a deep leg vein If part of the DVT breaks off it may lead to potentially life threatening complications such as pulmonary embolism If you have been inactive for prolonged periods and suspect you may have a DVT consult your doctor immediately To prevent dislodging the clot do not use Revitive In some cases of DVT there may be no symptoms However...

Страница 10: ...for your pain or suffer from severe pain such as burning pain in your legs and feet that continues even when you re resting You have suspected or diagnosed heart disease You have suspected or diagnosed epilepsy You are unsure about the suitability of the Revitive for you You are unsure about the cause of your symptoms If your conditions worsen If your condition worsen during or after use of the Re...

Страница 11: ...at balm gel moisturizer is evenly applied and thoroughly absorbed into the skin before applying stimulation There is a chance that uneven application of a balm gel moisturizer could increase the risk of skin irritation or burn when using the electrical stimulation Begin at the lowest intensity setting Place feet on the device and gradually begin to increase the intensity Never start at a high leve...

Страница 12: ...ath or shower while you are sleeping while you are driving operating machinery or during any activity in which electrical stimulation can put you at risk of injury In conjunction with a brace or cast without first consulting with your doctor Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation If such use is necessary thi...

Страница 13: ... period of immobility or inactivity use a low intensity and shorter time to avoid over stimulation or muscle fatigue This product is not intended for use by persons with reduced physical sensory or mental capabilities unless they are supervised by a person responsible for their safety Use this device only with the leads and accessories recommended by the manufacturer When using EMS It is possible ...

Страница 14: ...ffectiveness is highly dependent upon patient selection by a practitioner qualified in the management of pain patients The long term effects of electrical stimulation are unknown The safety of electrical stimulation during pregnancy has not been established Use caution following recent surgical procedures when stimulation may disrupt the patient s healing process Patients with suspected or diagnos...

Страница 15: ...date your warranty and may serious harm In the unlikely event your Revitive malfunctions disconnect it from the power source and contact your nearest authorized agent Revitive has passed the required tests for EMI it may still be affected by excessive emissions and or may interfere with more sensitive equipment It has been reported that some universal remote control devices e g for TV etc can chan...

Страница 16: ...n and burns beneath the electrodes have been reported with the use of powered muscle stimulator s You should stop using the device and should consult with your physician immediately if you experience adverse reactions from the device Save these instructions ...

Страница 17: ... 4 to help identify the parts b Locate the remote control and insert the batteries c Locate the AC DC power adapter Plug the small end of the power cord into the socket on the side of the Revitive d Plug the power adapter into the nearest electrical socket Before you use Revitive for the first time read the Important Safeguards on pages 9 16 If in doubt consult your doctor before using the product...

Страница 18: ...ortant to use Revitive at a high enough intensity to give you a strong comfortable muscle contraction Most people achieve a strong muscle contraction over intensity level 40 intensity range 1 99 If you are diagnosed with a long term medical condition such as Diabetes Osteoarthritis or a disease or condition associated with poor circulation such as for example diagnosed with varicose veins or Perip...

Страница 19: ...er button on the device or on the remote control to turn Revitive on You will hear a beep and the display will read 30 minutes To turn off Revitive at any time during operation press the power button on the device or on the remote control f Press the Intensity Control one time to start stimulation You can use the buttons on the control panel or remote control Ensure both feet are on the foot pads ...

Страница 20: ...tensity level 40 or over The IsoRocker enables Revitive to rock up and down once muscle contractions are nearing the desired level The rocking is caused by muscle contractions in the leg and the resulting ankle movement allowed by the IsoRocker is a good indicator that the intensity level is sufficient to provide the therapy Increasing the intensity above this level will enhance the therapy but sh...

Страница 21: ...ve cycles through its program Each time you adjust the intensity the intensity level shows in the display Again when you have finished adjusting the intensity the display reverts to the Minutes counter continuing to count down the session j When Revitive times out the Time Display reads 00 and you will hear three beeps k Revitive switches itself off automatically ...

Страница 22: ...the foot pads with a soft damp cloth Do not clean with chemicals Do not immerse Revitive in water c Store Revitive in a cool dry and dust free location Store out of direct sunlight Replacing the Batteries The remote control requires two AAA 1 5V batteries to operate Instructions for use ...

Страница 23: ...nd if your body is less hydrated below 60 then the stimulation may be less therefore it is important to always remain well hydrated The intensity level may be on too low a setting This is a very safe device Keep increasing the intensity level towards 99 until you feel the stimulation You may find that you have to increase the intensity level as you get used to the therapy The aim is not to get to ...

Страница 24: ...ity is adjusted to a level that is comfortable The device is too far in front of you Sit with your knees at a 90 degree angle When using the IsoRocker on a hard floor it makes a tapping noise Incorrect positioning of the device or too high an intensity level Adjust the positioning of the device or lower the intensity level to reduce the device tapping Alternatively use a floor mat under the device...

Страница 25: ...sical contact with the user SoftTouch Pads Foot Pads 195cm2 Durable Period Service Life of Device and Remote control 4 Years Power Source Power adapter UL Underwriters Laboratories Safety Mark Input Power adaptor used Output 100 240V AC 50 60Hz 0 18A 5 0V dc 1 0A The Remote Control replicates the controls found on the device Operating Frequencies 38 K Hz Operating Range Distance 0 to 25m at horizo...

Страница 26: ... 15 500Ω Foot 70mA 2kΩ Foot 48 5mA 10kΩ Foot 13 1mA Pulse Duration 10 Foot 450 970µs Frequency 10 Foot 20 to 44Hz Net Charge 500Ω 0 001 mC Maximum Phase Charge 500 Ω Foot 49uC Maximum Current Density Foot 0 026 mA cm Maximum Power Density 500Ω Foot 0 68 mW cm Burst Mode a Pulses per burst Foot 32 182 b Bursts per second Foot 0 1 0 3 c Burst duration Foot 1 90 8 4s d Duty Cycle Foot 0 56 0 89 ON Ti...

Страница 27: ...ain period of time contact the manufacturer Symbols Device serial number including year YYYY and month MM of manufacture can be found on the box and back of unit Item number Contraindications Warnings and Cautions Make sure you understand these before using Revitive Power Time Remaining Intensity Level Center Positive Polarity Class II medical electrical equipment double insulated Type BF medical ...

Страница 28: ...Y MM DD Humidity temperature and air pressure limit for storage and transport Humidity temperature and air pressure limit for operating conditions After any exposure to hot or cold temperatures outside the specified operating range of 50 104 F allow the product to re adjust to the recommended operating temperatures to ensure continued product performance Indoor Use Only Medical device does not con...

Страница 29: ...m the original purchase date Therefore it is vital to retain your original till receipt or invoice to indicate the date of initial purchase To qualify for the 2 year guarantee the device must have been used according to the manufacturer s instructions supplied Exclusions 1 Actegy manufacturer of Revitive devices shall not be liable to replace the goods under the terms of the guarantee where The fa...

Страница 30: ...30 ...

Страница 31: ...31 ...

Страница 32: ...d SoftTouch Pads are trademarks or registered trademarks of Actegy Ltd The Actegy product is a proprietary design and is protected by applicable design laws Legal Manufacturer Actegy Ltd 1 West Point Western Road Bracknell RG12 1HJ United Kingdom Tel 44 0 845 871 5989 Country Contact Distributor USA Actegy Health Inc 6400 Shafer Court Suite 225 Rosemont IL 60018 Toll free 1 877 260 3114 ...

Отзывы: