22
23
Output Specifications for Electrical Muscle Stimulation (EMS):
Waveform
Biphasic
Shape
Square symmetrical with polarity reversal
Maximum Output Voltage (+/-15%)
@500Ω
35Vp
Pulse Duration (+/-10%)
450-970µs
Net Charge @ 500Ω
[0.001]mC
Maximum Power Density @ 500Ω
0.68 mW/cm²
ON Time (+/-10%)
1.90 - 8.30s
OFF Time (+/-10%)
1.00 - 1.50s
The values of PULSE DURATIONS, PULSE repetition frequencies and amplitudes, including
any d.c. component, shall not deviate by more than ± 20 % when measured with a load
resistance within the range specified.
If confirmation is required that the Revitive works within its Essential Performance after a
certain period of time, contact the manufacturer
Complies with European Medical Devices Directive (93/42/EEC)
Device Lot number including year (YYYY) and month (MM) of
manufacture can be found on the box and back of unit
Item number
Contraindications, Warnings and Cautions
Make sure you understand these before using Revitive
Power
Time Remaining
Intensity Level
Center Positive Polarity
Class II medical electrical equipment double insulated
Type BF medical electrical equipment
Legal manufacturer of the device
EU/EC European Authorised Representative
Consult instructions for use
The Waste Electrical and Electronic Equipment Directive
(WEEE Directive).
At the end of the product lifecycle, do not throw this product
into normal household garbage, but take it to a collection point for the
recycling of electronic equipment
#YYYYMMXXXXX
LOT
REF
Symbols
Technical Specifications