SERENA Tecnico GB R-04 del 04.04.17
00/ Per Bertani / Ma nuale Tecnico Serena / Serena
Pagina 9
It is the max depression that the sterilizer must achieve in the vacuum phases. It is als o the point of alarm in
case the vacuum does not occur.
VACUUM MAX
0250 mbar
1.7 VACUUM DRY
CALIBRATION
INCREASE BAR
TIMED REFILLING
VACUUM MAX
VACUUM DRY
The values (D) indicate the range in which the bacteriological filter is activat ed
VACUUM DRY
MIN 0240 mbar
MAX 0400 mbar
1.8 VAPORIZER TEMPERATURE
TIMED REFILLING
VACUUM MAX
VACUUM DRY
VAPORIZER
TEMPERATURE
The following screen will appear:
The values indicate the range in which occurs the injection of water that becomes steam.
VAPORIZER
TEMPERATURE
MIN 150 °C
MAX 160 °C