SERENA Tecnico GB R-04 del 04.04.17
00/ Per Bertani / Ma nuale Tecnico Serena / Serena
Pagina 15
5
CALIBRATION AND ALIGNMENT
OPTIONS ->OPENING DOOR -> START
OPENING
DOOR
START
Once pressed START, use the arrows UP / DOWN to VISUALIZE THE TEMPERATURE
T1…c T2…c
Tv… c Tm… c
Tr… c Tc… c
Pr … mb Vo…im
…. Volt
…. Micron
A DF
Press INFO button to visualize the technician data with the door closed
T1 chamber probe temperature
T2 chamber probe temperature
Tv vaporizer probe temperature
Tm chamber metal temperature
Tr chamber resistance temperature
Tc condenser temperature
Pr chamber mb pressure
Vo water quantity volume
The data shown in the first line (eg +1.1 ° C) corresponds to the delta T of Calibration.
The data shown in the second row (eg 082.5 ° C) corresponds to the temperature detected at that time.