DIRECTIONS FOR USE
DISINFECTION SYSTEM
COMPOSITION :
The bath is equipped with an automatic system to determine the quantity of disinfectant.
The disinfection system includes :
- 1x 2 litres tank receiving the disinfectant product.
- 1X plunger with calibrated tube, located in the tank, and connected by a flexible going through the top of the venturi.
- 1 disinfection shower.
ADDITION OF DISINFECTANT :
Dilution :
The mixing of water and disinfectant inside the disinfection shower is made in 2 stages :
First :
Manual dilution inside the tank.
Second :
The suction device (venturi) ensures a dosage of 10% of the aspirated product from the tank.
Exemple :
The disinfectant BFR 4.0 must be used with a final dilution of 0,4%.
1)
Manual dilution inside the tank at 4%, that is 80ml of BFR 4.0 for 1920ml of water.
2)
Complementary dilution of 10% ensured with
cold water
by the suction device (venturi).
Once the diluted product is added to the tank, any unused solution will need to be discarded and refreshed after one week.
9/16