iBreeze CPAP Systems User Manual
Alert
Type
Possible Cause
Action
contact the supplier
if repeat.
therapy state; 2. Power
board short-circuit 3.
Humidifier heating
circuit short-circuit
The respiration
tube expired,
please replace.
Notification
In standby state, the
usage time exceed the
respiration tube setting
service life.
replace the
respiration tube.
The water tank
expired, please
replace.
Notification
In standby state, the
usage time exceed the
water tank setting
service life.
,
replace the water
tank.
The filter expired,
please replace.
Notification
In standby state, the
usage time exceed the
filter setting service
life.
Clic
replace the filter
The mask expired,
please replace.
Notification
In standby state, the
usage time exceed the
mask setting service
life.
replace the mask
It is time for device
maintenance,
Please contact the
service provider for
device
maintenance.
Notification
In standby state, the
usage time exceed the
setting device
maintenance time.
contact the service
provider for device
maintenance.
Low Minute
Volume (MV).
Alert 0
Minute volume less
than the setting
threshold value.
Check the respiration
tube or adjust therapy
parameter setting.
High Respiratory
Rate (RR).
Alert 0
Respiratory Rate
exceed the setting
threshold value.
Check the respiration
tube or adjust therapy
parameter setting.
Low Respiratory
Rate (RR).
Alert 0
None.
Check the respiration
tube or adjust therapy
parameter setting.
Apnea.
Alert 1
An apnea is detected
and exceeds the
setting duration.
Check the respiration
tube or adjust therapy
parameter setting.
High Leakage
Volume.
Alert 1
1. Inappropriate
connection of mask
and respiration tube.
Check the connection
of the mask or
respiration tube, and