Страница 2: ... The following symbols are used in this manual and on the inCourage Airway Clearance Therapy Their meanings are as follows RespirTech Business Hours Monday Friday 8am 5pm CST Toll Free 800 793 1261 www respirtech com 2896 Centre Pointe Drive St Paul MN 55113 TF 800 793 1261 Fax 800 962 1611 2014 Respiratory Technologies Inc 900000 000 Rev R General Warning Sign indicating the risk of personal inju...

Страница 3: ...Information Menu i 20 Retrieve Hour Meter Reading HMR 20 Retrieve Software Version 20 Retrieve Device Average Use Information 21 Setup Menu 21 Beep Control 22 Quick Start Auto Pause Session Time 22 Pause Time 22 Pressure Limit 23 Pressure Increment 23 Reset Defaults 23 Clear Average Use 23 Using the Button Lock 24 Cleaning and Filter Replacement 25 Troubleshooting Guide 27 Specifications 28 Contra...

Страница 4: ...to the chest to help loosen thin and move mucus throughout the lungs Indications for Use The inCourage Airway Clearance Therapy is indicated when external manipulation of the chest is the physician s treatment of choice for increasing the clearance of mucus in patients with pulmonary disorders The system promotes airway clearance and improves bronchial drainage utilizing High Frequency Chest Compr...

Страница 5: ...ir Interconnecting Hoses The two hoses are made of high strength fiber reinforced PVC The hoses deliver the pulses of air from the PTU to the vest Locking Connectors The connectors are made of nylon with stainless steel locking levers Power Cord The power cord plugs into the PTU and to a grounded three pronged outlet Grounding reliability can only be achieved when the equipment is connected to a p...

Страница 6: ...ged in and the On Off button is in the On position Stop Reset Button This button ends all current functions resets the control board and returns to the main menu QuickFit Adjustable Straps These straps allow you to size your vest to the exact repeatable recommended therapy spacing Air Ports The air ports are located at the front of the PTU and on the front of the vest The interconnecting hoses att...

Страница 7: ...alist at 1 800 793 1261 if any damage is suspected 7 Keep all components of the inCourage System away from heated surfaces to avoid damage and possible injury 8 Do NOT insert any object into any opening on the system 9 Always observe ALL warnings cautions and notes listed in this manual and posted on the inCourage System 10 Be sure to plug the inCourage System DIRECTLY into a properly grounded thr...

Страница 8: ...ps We have taken careful consideration in supplying you with the appropriate size vest based on your information Please take a moment and review the fitting instructions carefully and consult your physician if any discomfort develops NOTE We recommend a thin layer of cotton clothing to be worn under the vest for best comfort To loosen straps Lift the back of the buckle To tighten straps Hold the b...

Страница 9: ... bottom edge of the vest sits at the top of the hipbone 5 Gently tighten the front straps until each is snug to the chest 6 Pull each of the QuickFit blue tabs off the yellow tabs then place the blue tabs over the white tabs SYSTEM SETUP cont d 9 inCourage Airway Clearance Therapy INSTRUCTION MANUAL ...

Страница 10: ...ons 2 Plug the power cord to the back of the PTU 3 Plug the power cord into a properly grounded three pronged outlet Grounding reliability can only be achieved when the equipment is connected to a properly grounded three pronged outlet See product specifications for power requirements SYSTEM SETUP cont d 10 inCourage Airway Clearance Therapy INSTRUCTION MANUAL ...

Страница 11: ... metal lever firmly against the locking connector to secure to PTU Repeat steps 1 and 2 for the second hose 3 Lift the metal lever and place connector on the vest port as shown 4 Push the metal lever firmly against the locking connector to secure to vest Repeat steps 3 and 4 of these instructions for the second hose NOTE To disconnect the vest lift the metal lever and pull hose off the vest port S...

Страница 12: ...nnot be reset 8 USE INFO Represents the average session length in minutes and average pressure setting for the previous number of sessions displayed Only sessions over five minutes will be counted The user can reset this information 9 PAUSE TIME Allows for automatic continuation of therapy after a pre programmed pause during an Auto Pause or Multi Step session 10 PRESSURE LIMIT The maximum allowab...

Страница 13: ... Key Feature Program Session Length Pressure Frequency Memory Instant 30 minute1 ramping session Pre programmed 30 minute session divided into 12 steps of 2 5 minutes each1 Default set at 60 and adjustable during session Ramps from 6Hz to 15Hz over a 2 5 minute period then ramps from 15Hz to 6Hz over the next 2 5 minutes and repeats for a total of 30 minutes1 Last pressure setting used is retained...

Страница 14: ...anel will light green to show that the unit is in use 2 Adjust pressure by pressing either DOWN or UP Press PAUSE to temporarily stop the session Press RUN to resume the session or END to return to the startup screen STEP 2 Choose a Session Mode Quick Start Auto Pause Multi Step To run Quick Start mode 1 From the startup screen select QUICK START Your QUICK START session will begin A UT O M UL T I...

Страница 15: ...o temporarily stop the session or END to return to the startup screen 3 When the session pauses press RUN to resume the session or END to return to the startup screen 4 When a Pause Time is set a countdown to resume will be displayed and therapy will automatically resume when the counter reaches 0 00 The RUN button will also continue therapy at any time 1 8 9 6 0 2 9 3 4 P A US E DOW N U P E N D S...

Страница 16: ...m see the Modify a Multi Step Session section on page 17 4 The Program 1 session will begin Adjust pressure by pressing either DOWN or UP Press PAUSE to temporarily stop the session or END to return to the startup screen 5 If the session is paused press RUN to resume the session or END to return to the startup screen 6 When a Pause Time is set a countdown to resume will be displayed and therapy wi...

Страница 17: ...LT These options are described in the following instructions Fast Set Programming FS for Multi Step Sessions MULTI STEP PROGRAM EDIT FS PRESS Fast Set Programming creates a customized Multi Step session of up to 12 steps of alternating frequencies with an input of 6 settings Even Number of Steps Low Frequency odd steps High Frequency even steps Pressure Setting Total Program Time and Auto Pausing ...

Страница 18: ...rmation about the step The user can quickly check the settings of each step by pressing NEXT after reviewing a step The EXIT button will return to the Program Ready screen 4 Select EDIT to begin adjusting the parameters of the selected step that is currently displayed 5 Once in step EDIT mode pressing the BACK button at any time will return to the step summary for the previous parameter viewed 6 P...

Страница 19: ...ting DFLT as described in the Multi Step Editing Options section on page 17 will bring up the default confirmation screen To confirm program reset to default select YES Set a Multi Step Session to Factory Default Settings MULTI STEP PROGRAM EDIT DFLT PRESS Note As you complete programming for each step the display will briefly show that settings were saved C ONF I RM 1 T O DE F A UL T C ONF I RM Y...

Страница 20: ... END will return to the startup screen To Retrieve Hour Meter Reading HMR and Device Software Version 1 Press the Information i button to display the Information Menu 2 Press the HMR button 3 The HMR will be displayed on the right side of the screen Hours Minutes Seconds The software version will be displayed on the left side of the screen 4 Press the END button to return to the Information Menu T...

Страница 21: ...ational parameters that can be adjusted These parameters are the Beep Control QS AP Time Pause Time Pressure Limit Pressure Increment Default Settings and Clear Use Information The Setup Menu is displayed over two menu screens as shown to the right Selecting NEXT from Setup Menu 1 will display Setup Menu 2 To exit the Setup Menu select NEXT and or END to return to the Information Menu and then END...

Страница 22: ...minute increments If the session time is less than 30 minutes the number of 2 5 minute steps is reduced If session time is greater than 30 minutes the time is equally divided over 12 steps To adjust the Quick Start and Auto Pause session time select QS AP TIME from the Setup Menu 1 screen described on page 21 Select BACK to return to Setup Menu 1 without saving changes Pause Time When a Pause Time...

Страница 23: ... pressure increment to 5 or 10 When the desired setting is shown select SAVE Select BACK to return to Setup Menu 2 without saving changes Pressure Increment Reset Defaults and Clear Average Use This allows the user to reset settings to original factory parameters There are three reset options All defaults ALL DFLTS resets all programs and setup menu variables Setup defaults SETUP DFLTS resets only...

Страница 24: ...Lock 1 During any treatment session press and hold the PAUSE button 2 Within one second press and hold the END button 3 To disable the Button Lock repeat steps 1 and 2 NOTE The Button Lock will be disabled when the Reset button is pressed or when turning the power off K E Y B O ARD UNL OC KE D P A US E DOW N U P E ND K E Y B O ARD L OC KE D P A US E DOW N U P E ND 1 8 9 6 0 2 9 3 4 P A US E DOW N ...

Страница 25: ...est may be machine washed in any commercial or household washing machine by following these steps 1 Insert the two red plugs supplied into both vest ports Note Make sure the rim of the plug fits tightly into the port 2 Machine wash vest on gentle or normal cycle with a mild detergent in warm water 3 Remove both plugs from the vest ports and tumble dry vest on low heat setting until dry 4 To ensure...

Страница 26: ...ed that the filters be replaced monthly To replace the filter remove the two screws as shown below Remove and discard the old filter and slide the new filter into the filter holder Close filter door and replace screws WARNING Do not over tighten the screws NOTE If service is required please call a RespirTech Customer Care Representative at 1 800 793 1261 CLEANING and FILTER REPLACEMENT cont d 26 i...

Страница 27: ...t 1 800 793 1261 Re adjust the vest closures to accommodate change in chest size if the vest still does not fit properly contact the RespirTech Customer Care Representative at 1 800 793 1261 Connect hoses to the vest and PTU Clean inside and outside of the hoses vest ports and PTU ports according to maintenance guidelines Replace air filter Contact a RespirTech Customer Care Representative at 1 80...

Страница 28: ...ance Therapy Model ICS 1M US Manufactured By Mode of Operation Continuous Use Operating Temperature 10º C to 33º C 50º F to 92º F Ambient Temperature Storage Transport 25º C to 85º C 13º F to 185º F Operating Humidity 20 to 80 RH 25º C 77º F non condensing PTU Length 13 PTU Width 8 PTU Height 13 PTU Weight 17 lbs Input Source 120 VAC RMS 50 60Hz Single Phase Wattage 500 Watts Fuses 5 A 250 VAC 1 x...

Страница 29: ...pleural effusions or empyema Pulmonary embolism Rib fractures with or without flail chest Surgical wound or healing tissue recent skin grafts or flaps on the thorax Uncontrolled hypertension Distended abdomen Recent esophageal surgery Active or recent gross hemoptysis Uncontrolled airway at risk for aspiration tube feeding or recent meal Subcutaneous emphysema Recent epidural spinal infusion or sp...

Страница 30: ...romagnetic environment specified below The user of the inCourage System should assure that it is used in such an environment The inCourage System Model ICS 1M US uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment The inCourage System Model ICS 1M US is suitable for use in all establishmen...

Страница 31: ...ent If the user of the inCourage System requires continued operation during power mains interruptions it is recommended that the inCourage System be powered from an uninterruptible power supply or a battery Power frequency 50 60 Hz magnetic field IEC 61000 4 8 3 A m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environm...

Страница 32: ...th care providers that is sufficient to assure access to appropriate high quality health care including giving women access to qualified specialists such as obstetrician gynecologists and giving patients with serious medical conditions and chronic illnesses access to specialists 4 Access to Emergency Services Patients have the right to access emergency health services when and where the need arise...

Страница 33: ...the company address Respiratory Technologies Inc Attn Privacy Official 2896 Centre Pointe Drive St Paul MN 55113 A Company representative will respond back to the patient within five 5 days to confirm receipt of the notification of concern and will initiate an investigation of the alleged grievance or abuse concern FOR MEDICARE BENEFICIARIES Within fourteen 14 days of receipt of the beneficiary s ...

Страница 34: ... to you All RespirTech staff members have been trained to protect privacy Staff members who violate these policies are subject to disciplinary action We use safeguards to protect your health data These safeguards comply with federal regulations regarding security We periodically review our policies and practices We monitor our computer networks We monitor and test our security to ensure the privac...

Страница 35: ...t notifying you of such uses or disclosures These circumstances may include when the use or disclosure is For public health activities For example when reporting to public health authorities the exposure to certain diseases or reporting data about immunizations For health oversight activities For example when disclosing health data to a state or federal health oversight agency so it can monitor th...

Страница 36: ...h data You need to make your request in writing and explain your reason for the amendment Write or call RespirTech and ask for a RespirTech Information Amendment Request Form Under limited circumstances we may deny your request If we do so you may file a statement of disagreement with us You may also ask that any future disclosures of your health data include your requested amendment and our denia...

Страница 37: ...ogies Inc Attn Privacy Official 2896 Centre Pointe Drive St Paul MN 55113 1 800 793 1261 Toll Free or 651 379 8960 Direct Dial privacyofficial respirtech com You may also submit a written complaint to the U S Department of Health and Human Services at Region V Office for Civil Rights U S Department of Health and Human Services 233 N Michigan Avenue Suite 240 Chicago IL 60601 Hotline 800 368 1019 W...

Страница 38: ...A supplier must notify beneficiaries of warranty coverage and honor all warranties under applicable State law and repair or replace free of charge Medicare covered items that are under warranty 7 A supplier must maintain a physical facility on an appropriate site This standard requires that the location is accessible to the public and staffed during posted hours of business The location must be at...

Страница 39: ...l or control interest in the supplier 18 A supplier must not convey or reassign a supplier number i e the supplier may not sell or allow another entity to use its Medicare billing number 19 A supplier must have a complaint resolution protocol established to address beneficiary complaints that relate to these standards A record of these complaints must be maintained at the physical facility 20 Comp...

Страница 40: ...2 C F R 424 57 c Implementation date May 4 2009 27 A supplier must obtain oxygen from a state licensed oxygen supplier 28 A supplier must maintain ordering and referring documentation consistent with provisions found in 42 C F R 424 516 f 29 DMEPOS suppliers are prohibited from sharing a practice location with certain other Medicare providers and suppliers 30 DMEPOS suppliers must remain open to t...