Chapter 4
4-10
BiPAP Vision Service Manual © Respironics, Inc.
1062788
Periodic Maintenance
4.6.3
Verifying the Charge
At least two hours is required to charge a fully discharged battery to a voltage
to a level that does not activate a low voltage alarm. After 2 hours, the
ventilator can be operated and continues to trickle-charge the battery during
operation.
1. Press Monitoring to begin operation.
2. Wait 2 minutes to determine if the
Check Vent
alarm activates and an
error code 205 is displayed in the Error Message screen. If not, then
the unit is ready for use.
Содержание BiPAP Vision
Страница 1: ...BiPAP Vision Ventilatory Support System S E R V I C E M A N U A L ...
Страница 120: ...Chapter 6 6 28 BiPAP Vision Service Manual Respironics Inc 1062788 Troubleshooting This page is intentionally blank ...
Страница 212: ...Appendix A A 42 BiPAP Vision Service Manual Respironics Inc 1062788 This page is intentionally blank ...
Страница 221: ...Appendix C 1062788 BiPAP Vision Service Manual Respironics Inc C 3 Figure C 2 MC Board Analog Outputs ...
Страница 222: ...Appendix C C 4 BiPAP Vision Service Manual Respironics Inc 1062788 Figure C 3 MC Board Watchdog Real Time Clock ...
Страница 224: ...Appendix C C 6 BiPAP Vision Service Manual Respironics Inc 1062788 Figure C 5 MC Board Power Supply Status ...
Страница 225: ...Appendix C 1062788 BiPAP Vision Service Manual Respironics Inc C 7 Figure C 6 MC Board Bypass Circuitry ...
Страница 227: ...Appendix C 1062788 BiPAP Vision Service Manual Respironics Inc C 9 Figure C 8 DC Battery and Indicator Circuitry ...
Страница 228: ...Appendix C C 10 BiPAP Vision Service Manual Respironics Inc 1062788 Figure C 9 DC ELC Circuitry ...
Страница 231: ...Appendix C 1062788 BiPAP Vision Service Manual Respironics Inc C 13 Figure C 12 DC Microcontroller Interface ...
Страница 232: ...Appendix C C 14 BiPAP Vision Service Manual Respironics Inc 1062788 Figure C 13 DC Board Power Conversion Circuitry ...
Страница 234: ...Appendix C C 16 BiPAP Vision Service Manual Respironics Inc 1062788 Figure C 15 DC Board ELC Circuitry ...
Страница 236: ...Appendix C C 18 BiPAP Vision Service Manual Respironics Inc 1062788 Figure C 17 DC Board Microcontroller Interface ...
Страница 237: ...Appendix C 1062788 BiPAP Vision Service Manual Respironics Inc C 19 Figure C 18 Pressure Control PC Board CPU ...
Страница 238: ...Appendix C C 20 BiPAP Vision Service Manual Respironics Inc 1062788 Figure C 19 PC Board Motor Controller ...
Страница 239: ...Appendix C 1062788 BiPAP Vision Service Manual Respironics Inc C 21 Figure C 20 PC Board Pressure Sensors ...
Страница 240: ...Appendix C C 22 BiPAP Vision Service Manual Respironics Inc 1062788 Figure C 21 PC Board Valve Drivers ELC ...
Страница 241: ...Appendix C 1062788 BiPAP Vision Service Manual Respironics Inc C 23 Figure C 22 PC Board Bypass Pullup ICB Databus ...
Страница 242: ...Appendix C C 24 BiPAP Vision Service Manual Respironics Inc 1062788 Figure C 23 Air Flow Module AFM ...
Страница 243: ...Appendix C 1062788 BiPAP Vision Service Manual Respironics Inc C 25 Figure C 24 Oxygen Module OM ...
Страница 244: ...Appendix C C 26 BiPAP Vision Service Manual Respironics Inc 1062788 Figure C 25 Power Supply Subsystem PSS ...