Chapter 7
7-36
BiPAP Vision Service Manual © Respironics, Inc.
1062788
Testing and Calibration
7.8 Oxygen Module
(OM) Calibration
The OM calibration adjusts the zero voltage on the OM assembly main board.
Perform this calibration on any OM module that is suspected to be out of
tolerance, then perform the complete system final test on the ventilator.
NOTE:
Perform OM calibration on the existing OM before replacing any of its
components. If it is necessary to replace an OM component, test the OM
using the system final test (Section 7.7). If OM testing fails, perform OM
calibration then repeat the system final test.
NOTE:
If the OM still does not pass the system final test, follow standard repair
procedures.
Required equipment: Fluke 87 digital multimeter (DMM) or equivalent (3 1/2-
digit readout)
1. Remove the complete OM assembly except for its ground wire and
ribbon cable, and set next to the ventilator.
2. Connect the DMM to TP9 and TP13 (ground).
3. Power on the ventilator and allow it to complete its system self test.
Do
not
enter Monitoring.
4. Adjust R41 on the OM so that the DMM measures 0.225-0.235 VDC.
5. Reassemble the ventilator.
Содержание BiPAP Vision
Страница 1: ...BiPAP Vision Ventilatory Support System S E R V I C E M A N U A L ...
Страница 120: ...Chapter 6 6 28 BiPAP Vision Service Manual Respironics Inc 1062788 Troubleshooting This page is intentionally blank ...
Страница 212: ...Appendix A A 42 BiPAP Vision Service Manual Respironics Inc 1062788 This page is intentionally blank ...
Страница 221: ...Appendix C 1062788 BiPAP Vision Service Manual Respironics Inc C 3 Figure C 2 MC Board Analog Outputs ...
Страница 222: ...Appendix C C 4 BiPAP Vision Service Manual Respironics Inc 1062788 Figure C 3 MC Board Watchdog Real Time Clock ...
Страница 224: ...Appendix C C 6 BiPAP Vision Service Manual Respironics Inc 1062788 Figure C 5 MC Board Power Supply Status ...
Страница 225: ...Appendix C 1062788 BiPAP Vision Service Manual Respironics Inc C 7 Figure C 6 MC Board Bypass Circuitry ...
Страница 227: ...Appendix C 1062788 BiPAP Vision Service Manual Respironics Inc C 9 Figure C 8 DC Battery and Indicator Circuitry ...
Страница 228: ...Appendix C C 10 BiPAP Vision Service Manual Respironics Inc 1062788 Figure C 9 DC ELC Circuitry ...
Страница 231: ...Appendix C 1062788 BiPAP Vision Service Manual Respironics Inc C 13 Figure C 12 DC Microcontroller Interface ...
Страница 232: ...Appendix C C 14 BiPAP Vision Service Manual Respironics Inc 1062788 Figure C 13 DC Board Power Conversion Circuitry ...
Страница 234: ...Appendix C C 16 BiPAP Vision Service Manual Respironics Inc 1062788 Figure C 15 DC Board ELC Circuitry ...
Страница 236: ...Appendix C C 18 BiPAP Vision Service Manual Respironics Inc 1062788 Figure C 17 DC Board Microcontroller Interface ...
Страница 237: ...Appendix C 1062788 BiPAP Vision Service Manual Respironics Inc C 19 Figure C 18 Pressure Control PC Board CPU ...
Страница 238: ...Appendix C C 20 BiPAP Vision Service Manual Respironics Inc 1062788 Figure C 19 PC Board Motor Controller ...
Страница 239: ...Appendix C 1062788 BiPAP Vision Service Manual Respironics Inc C 21 Figure C 20 PC Board Pressure Sensors ...
Страница 240: ...Appendix C C 22 BiPAP Vision Service Manual Respironics Inc 1062788 Figure C 21 PC Board Valve Drivers ELC ...
Страница 241: ...Appendix C 1062788 BiPAP Vision Service Manual Respironics Inc C 23 Figure C 22 PC Board Bypass Pullup ICB Databus ...
Страница 242: ...Appendix C C 24 BiPAP Vision Service Manual Respironics Inc 1062788 Figure C 23 Air Flow Module AFM ...
Страница 243: ...Appendix C 1062788 BiPAP Vision Service Manual Respironics Inc C 25 Figure C 24 Oxygen Module OM ...
Страница 244: ...Appendix C C 26 BiPAP Vision Service Manual Respironics Inc 1062788 Figure C 25 Power Supply Subsystem PSS ...