7. The dressing assistant should now switch off the AFU and unscrew the
externally mounted filters for safe disposal (re-seal filters with the original
plastic caps).
8. If necessary the peak-less safety helmet can now be recovered from the head
space of the suit.
9. The suit, including attached AFU, can now be sealed in the hazbag using the
plastic tie provided (fig. 36).
If required enter all relevant information onto
the hazbag tag
.
Emergency self doffing procedure
1. Withdraw your right arm from the suit sleeve and push against the back of the
zip at the pull tag end.
2. With your left hand unseal the outer velcro flap fastening and undo the zip.
3. Undo the AFU belt and unfasten the helmet chin strap (if worn).
4. Exit the suit.
Rapid self doffing may also be achieved by cutting the suit (this will render the suit
unsuitable for any future use).