background image

46

47

Reference test gain (60 dB SPL input)

HFA

31

dB

Full-on gain (50 dB SPL Input)

Max
HFA

52
43

dB 
dB

Maximum output (90 dB SPL input)

Max
HFA

113

108

dB SPL
dB SPL

Total harmonic distortion

500 Hz
800 Hz

1600 Hz

0.3
0.5
0.7

%
%
%

Telecoil sensitivity (SPLIV @ 31,6 mA/m)

90

dB SPL

Equivalent input noise (w/o noise reduction)

23

dB SPL

Frequency range (DIN 45605)

100–
7060

Hz

Current drain (in test mode)

1.3

mA

RIE - LP receiver

Models: LS962-DRW, LS762-DRW, LS562-DRW
LS961-DRW, LS761-DRW, LS561-DRW

Frequency (Hz)

FULL-ON AND REFERENCE TEST GAIN

FULL-ON AND REFERENCE TEST GAIN

G

ain

 

(d

B

)

Reference test gain 

60 dB SPL input

Full-on gain 

50 dB SPL input

2cc Coupler

MAXIMUM OUTPUT (OSPL 90) 

MAXIMUM OUTPUT (OSPL 90) 

Frequency (Hz)

O

ut

p

ut

 

(d

B

 S

P

L)

2cc Coupler

Data in accordance with ANSI S3.22–2009 and IEC 60118-7; 

Supply Voltage 1.3 V, 2cc coupler

Note: Telecoil sensitivity only applicable for 62 RIE models

Note: Telecoil sensitivity only applicable for 62 RIE models

Reference test gain (60 dB SPL input)

HFA

37

dB

Full-on gain (50 dB SPL Input)

Max
HFA

58
51

dB
dB

Maximum output (90 dB SPL input)

Max
HFA

116
114

dB SPL
dB SPL

Total harmonic distortion

500 Hz
800 Hz

1600 Hz

0.5
0.6
1.2

%
%
%

Telecoil sensitivity (SPLIV @ 31.6 mA/m)

96

dB SPL

Equivalent input noise (w/o noise reduction)

23

dB SPL

Frequency range (DIN 45605)

100–
7000

Hz

Current drain (in test mode)

1.3

mA

RIE—MP receiver

Models: LS962-DRW, LS762-DRW, LS562-DRW
LS961-DRW, LS761-DRW, LS561-DRW

Frequency (Hz)

G

ain

(d

B

)

Reference test gain

60 dB SPL input

Full-on gain

50 dB SPL input

2cc Coupler

Frequency (Hz)

O

ut

p

ut

(d

B

 S

P

L)

2cc Coupler

Data in accordance with ANSI S3.22–2009 and IEC 60118-7; 

Supply Voltage 1.3 V, 2cc coupler

Technical data

Содержание BRIE

Страница 1: ...USER GUIDE RECEIVER IN EAR HEARING INSTRUMENT...

Страница 2: ...the receiving antenna Increase the separation between the equipment and receiver Connect the equipment into an outlet on a circuit different from the one in which the receiver is con nected Consult t...

Страница 3: ...d permanent hearing damages the maximum daily usage depends on the level of the generated sound Should you develop any side effects from using the sound generator such as dizziness nausea head aches p...

Страница 4: ...stru ments of type BRIE with FCC ID X26BRIE IC number 6941 BRIE and size 13 battery are available in the following variants LS962 DRW LS762 DRW LS562 DRW The identification number for the BRIE and VE3...

Страница 5: ...udio boots 25 Care and maintenance 26 Daily maintenance 27 The receiver tube 28 Cleaning the receiver tubes and domes 28 Cleaning RIE or moulds 28 Changing wax guard for receiver tube 29 How to apply...

Страница 6: ...Tulip Dome 4 Receiver Power Dome 5 RIE mould 6 Sports lock 7 Battery compartment 8 Receiver 9 Push button 10 Volume Control optional 11 LP receiver tube 12 MP receiver tube 13 HP receiver tube 14 Mod...

Страница 7: ...usual but you will not be able to use your ReSound Unite TV Streamer 2 Phone Clip and Mini Microphone At some point the battery level will not support the remote control either and you will once again...

Страница 8: ...s removing the hearing instrument Sports lock The sports lock will be applied or adjusted by your hearing care professional Inserting Removing hearing instruments Insertion custom RIE and UP moulds 1...

Страница 9: ...in programme 2 it will switch to programme 3 etc When you close the battery door and switch the instrument on it will start in programme 1 Press the pro gramme or multi function buttons if you want t...

Страница 10: ...function resumes before open ing and closing the battery compartment again for any reason If the battery compartment is opened and closed during this 15 second window flight mode will resume If you h...

Страница 11: ...out by listening to news commentators since they usually speak clearly then try other programmes If you find it difficult to listen to TV or radio your hearing care professional will be able to give...

Страница 12: ...ne receiver slightly to find the best position for reliable PhoneNow activation and good hearing on the telephone If you have two hearing instruments with the synchronization function enabled the volu...

Страница 13: ...system may not be turned on or is not operating correctly If a facility is not equipped with a tele loop system sitting as close as possible to the front may be helpful Direct audio input optional Us...

Страница 14: ...5 Do not wear your instrument while showering swimming in heavy rain or in a moist atmosphere such as a steam bath or sauna 6 If your instrument does get wet or if it has been exposed to high humidit...

Страница 15: ...ng the receiver tubes or the receiver domes This process is also used to clean the UP receiver mould Please see instruction on page 29 for how to change the wax guard filter iCleaning RIE or moulds do...

Страница 16: ...of the smaller petal Using ReSound hearing instruments with smart phone apps iIntended use of smart phone apps GN ReSound smart phone apps are intended to be used with GN ReSound wireless hearing aid...

Страница 17: ...ment mobile telephones contain less energy and will not damage hearing instruments However they have the potential to momentarily affect the sound quality or temporarily create strange sounds from hea...

Страница 18: ...specific therapeutic needs and your personal preference as determined by your doctor audiologist or hearing care professional Depending on the selected hearing instrument programme and the environmen...

Страница 19: ...epth of up to 14dB High pass filter Low pass filter 500 Hz 2000 Hz 750 Hz 3000 Hz 1000 Hz 4000 Hz 1500 Hz 5000 Hz 2000 Hz 6000 Hz iTSG warnings Sound generators can be dangerous if improperly used Sou...

Страница 20: ...level the device should not be used for more than two 2 hours per day In no case should the sound generator be worn at uncomfortable levels iBattery warning information Batteries although very small c...

Страница 21: ...medical evaluation the physician will give you a written statement that states that your hear ing loss has been medically evaluated and that you may be considered a candidate for a hearing aid The ph...

Страница 22: ...itivity Excessive ear wax Volume set too low POSSIBLE REMEDY Turn on by closing the battery door Replace battery Insert battery properly Clean RIE mould or tube Replace wax filter or consult your hear...

Страница 23: ...s in flight mode POSSIBLE REMEDY Re insert RIE mould carefully Re insert dome Consult your hearing care professional Consult your hearing care professional Consult your hearing care professional Consu...

Страница 24: ...r Data in accordance with ANSI S3 22 2009 and IEC 60118 7 Supply Voltage 1 3 V 2cc coupler Note Telecoil sensitivity only applicable for 62 RIE models Note Telecoil sensitivity only applicable for 62...

Страница 25: ...st gain 60 dB SPL input HFA 47 dB Full on gain 50 dB SPL Input Max HFA 75 64 dB dB Maximum output 90 dB SPL input Max HFA 129 124 dB SPL dB SPL Total harmonic distortion 500 Hz 800 Hz 1600 Hz 1 3 2 1...

Страница 26: ...etween 25 C and 70 C according to internal and industry standards During transport or storage the temperature should not exceed the limit values of 20 C to 60 C and relative humidity of 90 RH non cond...

Страница 27: ...ralia GN ReSound Pty Ltd Unit R1 Regents Park Estate 391 Park Road Regents Park NSW 2143 Tel free 1800 658 955 Fax 61 2 9743 7472 www gnresound com au New Zealand GN ReSound NZ Ltd 12 Parkway Drive Ma...

Отзывы: