5 Replacing major subassemblies
27
6 Fit the new blower inlet connector, ensuring it is seated correctly on the volute.
7 Before placing the volute in the casing, check that the motor cable is located under the tab, as
shown below:
8 Place the volute in the blower casing. Ensure the blower inlet and outlet connectors are seated
as shown below:
9 Place the motor cable grommet in the slot as shown below:
Blower inlet connector
flush to wall of volute
Motor cable located
under tab
Grommet in slot in
blower casing