3 General inspection and servicing
9
1111
Clean the flow generator
Clean the flow generator
Clean the flow generator
Clean the flow generator
Before servicing the unit it should be wiped down using a lint free cloth and a cleaning agent with a
high alcohol content, for example, isopropyl alcohol.
Caution
When using a cleaning agent with a high alcohol content refer to the appropriate Material
Safety Data Sheet to ensure safe handling of the cleaning agent.
Remove any stains from the cases. When removing any stains from the flow generator or mask
system use soap or detergent-based cleaning solutions. Remove excess and wipe dry after use.
Do
not
use thinners, trichlorethylene, bleach or other solvents.
Do not
apply oil to any component
of the flow generator.
2222
Record clinical settings
Record clinical settings
Record clinical settings
Record clinical settings
1 Open AutoSet Spirit Service Software 2.1.
2 Select
Settings>Flow generator settings
from the main window, or press the
Open flow
generator settings dialog
button at the top of the main window.
3 The following screen will appear with the flow generator settings:
4 Select
File>Save screen settings to file
, or select the floppy disk icon and save the settings
under the unit’s serial number.
Warning
If the clinical settings can not be recorded, press the
Default
button and save the file with
the default settings for the faulty unit. Notify the patient that the unit has been reset to
default settings.
!
Main window in service software
!