31
Cleaning and Maintenance
Cleaning and Maintenance
You should regularly carry out the cleaning and maintenance described in this
manual.
!
CAUTION
Do not wash the pressure sensor tube. If fluid enters the pressure sensor
tube allow it to dry completely by hanging it in a clean place out of direct
sunlight. If the pressure sensor tube cannot be dried completely it should be
replaced.
Daily
1. Disconnect the air tubing and pressure sensor tube and hang them in a clean,
dry place until next use. Do not hang the air tubing in direct sunlight as it may
harden and crack over time.
2. Clean the mask according to the mask user instructions.
3. If you are using a humidifier, clean it according to the instructions in the
manual.
Weekly
1. Remove the air tubing and pressure sensor tube from the
VPAP Adapt
SV
unit
and
the mask.
2. Remove the pressure sensor tube and proximal cuff from the air tubing.
3. Wash the mask system according to the instructions supplied with it.
4. Wash the air tubing and Proximal cuff in warm water using mild liquid soap.
Rinse thoroughly, hang, and allow to dry.
5. Before next use, assemble the mask and headgear according to the mask
user instructions.
6. Reconnect the pressure sensor tube to the air tubing, then connect both to
the Proximal cuff. Attach the Proximal cuff to the mask.
Содержание VPAP Adapt SV
Страница 1: ...VPAP Adapt SV User Guide English Front Cover User qxp 05 04 2006 4 31 PM Page 1...
Страница 3: ...VPAP Adapt SV User Guide English Reorder number 26869 2 06 04...
Страница 10: ...4...
Страница 12: ...6...
Страница 32: ...26...
Страница 36: ...30...
Страница 42: ...36...
Страница 44: ...38...
Страница 54: ...48...
Страница 56: ...50 System Error 7 message 35 T treatment screens 25 U user owner responsibility 1 V VPAP Adapt SV 7 usage 24...