Cleaning and maintenance
The Remote Alarm can be cleaned using an anti-bacterial
cleaning solution such as Mikrozid AF
®
or CaviCide
®
on a clean,
non-dyed disposable cloth. For a full list of approved cleaning
solutions, refer to the Technical Specifications section.
Troubleshooting
Problem/possible cause
Solution
The LED does not turn on when connected to the ventilator or
when the Self-test button is pressed.
Batteries may not be
inserted or may be flat
Check that two AA batteries are inserted
correctly in the rear of the alarm. If
necessary, change the batteries. If the
problem persists, contact your ResMed
service agent.
The Remote Alarm detects a disconnect while the cable is
connected.
Poor connection
Change the cable. If the problem persists,
change the Remote Alarm.
The Remote Alarm detects a disconnect while an alarm is raised on
the ventilator.
Poor connection
Change the cable. If the problem persists,
change the Remote Alarm.
The second Remote Alarm is sounding an alarm with no alarm
raised on the ventilator.
Poor connection.
Check that the connecting cable is firmly
plugged into the first Remote Alarm and
the rear of the ventilator. If the problem
persists, contact your ResMed service
agent.
The Remote Alarm triggers a low battery alarm.
Batteries are flat.
Change the batteries.
Technical specifications
Operating temperature
0°C to +40°C
Operating humidity
10% to 95% non-condensing
Operating altitude
Sea level to 3,050 m
Storage and transport
temperature
-30°C to +70°C
Storage and transport
humidity
10% to 95% non-condensing
Electromagnetic compatibility
This product complies with
all applicable electromagnetic
compatibility requirements (EMC)
according to IEC60601-1-2:2007,
for residential, commercial and
light industry environments The
electromagnetic compatibility tables
for these ResMed devices can be
found on www.resmed.com, on
the Products page under Service &
Support. Click on the PDF file for
your device.
Output volume
55–90 dBa (as per ISO 3744:2010)
Содержание Power Station II
Страница 2: ...10 9 8 7 1 2 3 4 6 5 A B C...
Страница 99: ...97 4 LR6 12 C 1 3 2 5 LED 10 3 1 2 3 D 1 2 5 3 3 4 5...
Страница 100: ...98 1 2 30 m 5 3 4 30 m 1 2 3 Low Battery Disconnect...
Страница 101: ...99 LED LED LED LED LED 10 LED 4 LED 10 T...
Страница 102: ...100 LED LED LED LED LED LED LED 10 LED...
Страница 103: ...101 1 2 3 Mikrozid AF CaviCide LED AA ResMed ResMed 0 C 40 C 10 95 3 050 m 30 C 70 C...
Страница 105: ...103 IP22 II RoHS CSA LOT REF WEEE 2002 96 EC ResMed ResMed www resmed com environment ResMed ResMed ResMed...
Страница 123: ...121 II ResMed Astral Stellar ResMed A 1 6 3 2 7 3 8 4 9 5 5 10...
Страница 125: ...123 3 D 1 2 5 3 3 4 5 1 2 30 5 3 4 30...
Страница 126: ...124 1 2 3 10 4...
Страница 127: ...125 10...
Страница 128: ...126 10 E 1 2 3 Mikrozid AF CaviCide...
Страница 148: ...146 A 1 6 3 2 7 3 8 4 9 5 5 10 II 2 5 10 20 30 www resmed com B 1 2 AA1 5 3 4 LR6 AA 12 C 1 3 2 5 LED 10 1 3 1 2 3...
Страница 149: ...147 D 1 2 5 3 3 4 5 1 2 30 5 3 4 30 1 2 3...
Страница 150: ...148 LED LED LED LED LED 10 LED 4 LED 10 1 LED LED LED LED...
Страница 151: ...149 LED 10 LED E 1 2 3 Mikrozid AF CaviCide LED AA...
Страница 153: ...151 WEEE 2002 96 EC X www resmed com environment 12 a b c d II Astral Stellar...
Страница 154: ...152 A 1 6 3 2 7 3 8 4 9 5 5 10 II 2 5 10 20 30 www resmed com B 1 2 AA1 5 3 4 LR6 AA 12 C 1 3 2 5 LED 10 1 3...
Страница 155: ...153 1 2 3 D 1 2 5 3 3 4 5 1 2 30 5 3 4 30 1 2 3...
Страница 156: ...154 LED LED LED LED LED 10 LED 4 LED 10 1...
Страница 157: ...155 LED LED LED LED LED 10 LED E 1 2 3 Mikrozid AF CaviCide LED AA...
Страница 159: ...157 IP22 II RoHS LOT REF WEEE 2002 96 EC X www resmed com environment 12 a b c d...
Страница 160: ...D E...