8
Technical specifications
Therapy
pressure
2 to 30 cm H
2
O
Compliance
Mask compliance between 5 cm H
2
O and 15 cm H
2
O is less than
1.57 mL/cm H
2
O
Resistance
Drop in pressure measured (nominal)
at 50 L/min: 0.8 cm H
2
O
at 100 L/min: 3.1 cm H
2
O
Sound
DECLARED DUAL-NUMBER NOISE EMISSION VALUES in
accordance with ISO 4871. The A-weighted sound power level of the
mask is 40.6 dBA, with uncertainty 4 dBA. The A-weighted sound
pressure level of the mask at a distance of 1 m is 32.6 dBA, with
uncertainty 4 dBA.
Environmental
conditions
Operating temperature: +5°C to +40°C
Operating humidity: 15% to 95% non-condensing.
Storage and transport: -20°C to +60°C
Storage and transport humidity: up to 95% non-condensing.
Gross
dimensions
Mask fully assembled with elbow assembly – no headgear.
35 mm (H) x 430 mm (W) x 40 mm (D)
Notes:
•
The mask system does not contain PVC, DEHP or phthalates.
•
This product does not contain natural rubber latex.
•
The manufacturer reserves the right to change these specifications without notice.
Storage
Ensure that the mask is thoroughly clean and dry before storing it for any length of time.
Store the mask in a dry place out of direct sunlight.
Disposal
This mask does not contain any hazardous substances and may be disposed of with your
normal household refuse.
Symbols
Caution, consult accompanying documents;
LOT
Batch Code;
REF
Catalogue number;
Temperature limitation;
Humidity limitation;
Does not contain natural rubber latex;
Manufacturer;
European Authorised
Representative;
Indicates a Warning or Caution and alerts you to a possible injury or
explains special measures for the safe and effective use of the device;
Keep away
from rain;
This way up; Fragile, handle with care.
Consumer warranty
ResMed acknowledges all consumer rights granted under the EU Directive 1999/44/EG
and the respective national laws within the EU for products sold within the European
Union.
618216r2 Pixi Pediatric User Guide EPN.indb 8
22/05/2012 4:16:19 PM