Ventilation modes
70
Parameter
Setting
Trigger
When Trigger Type is set to
Flow
:
Adult: Off, 0.5 to 15 [1.0] (L/min)
Paed: Off, 0.5 to 15 [0.5] (L/min)
When Trigger Type is set to
Pressure
:
Off, Very Low to Very High [Medium]
Rise Time
(msec)
Min, 150 to 900 [200]
Note: Some default settings are different for Mouthpiece circuit.
Supplementary features:
•
Safety Vt (Tidal Volume)
•
Apnoea response
•
Sigh breath (Astral 150 only)
•
Manual breath (Astral 150 only)
The P(A)CV breath pattern showing one patient triggered breath amongst time-triggered breaths. The
Safety Vt feature is turned off.