Troubleshooting
English 161
Error code
Action
204
Unable to learn the circuit.
1.
Ensure that the circuit is not moved until completion of the test.
2.
Check the circuit and attached accessories for blockages.
3.
Ensure that there are no sharp bends or kinks in the circuit and the patient end is not
blocked.
4.
If using humidification, ensure that the humidifier tub is not overfilled.
5.
Follow the on-screen instructions carefully:
•
circuit should not be blocked during step 2
•
circuit should be completely blocked during step 3.
If problem persists, this circuit may not be compatible with the Astral device. Consider
selecting another circuit configuration.
205
The measured circuit resistance exceeds safe operating limits for this device.
1.
Check the circuit and attached accessories for blockages.
2.
Ensure that there are no sharp bends or kinks in the circuit and the patient end is not
blocked.
3.
If using humidification, ensure that the humidifier tub is not overfilled.
4.
Follow the on-screen instructions carefully:
•
circuit should not be blocked during step 2
•
circuit should be completely blocked during step 3.
If problem persists, this circuit may not be compatible with the Astral device. Consider
selecting another circuit configuration.
206
Hardware fault detected. Contact an authorised Service Centre.
303
Unable to calibrate oxygen sensor.
1.
Check that supplemental oxygen is not added during the Learn Circuit.
2.
Repeat Learn Circuit. If the problem persists, replace the oxygen sensor as described
in Replacing the Oxygen sensor.
404, 405, 406
Hardware fault detected. Contact an authorised Service Centre.
409
Learn Circuit was unable to complete due to excessive leak from the circuit.
1.
Check that the circuit is completely blocked during the third step of the Learn Circuit.
2.
Check that the circuit is assembled correctly and there are no leaks in the circuit.
3.
Check that the circuit adapter is firmly inserted.
4.
This circuit may not be compatible with the Astral device. Try another circuit.
Repeat Learn Circuit. If the problem persists, contact an authorised Service Centre.
415
Hardware fault detected. Contact an authorised Service Centre.
420
The measured circuit compliance exceeds safe operating limits for this device.
Check that the circuit is assembled correctly and completely blocked during the third step
of the Learn Circuit.
If problem persists, this circuit may not be compatible with the Astral device. Consider
selecting another circuit configuration.
426
Hardware fault detected. Contact an authorised Service Centre.