Air tubing
Air tubing
Material
Length
Inner diameter
ClimateLineAir
Flexible plastic and electrical components
6'6" (2 m)
0.6" (15 mm)
ClimateLineAir Oxy
Flexible plastic and electrical components
6'4" (1.9 m)
0.75" (19 mm)
SlimLine
Flexible plastic
6' (1.8 m)
0.6" (15 mm)
Standard
Flexible plastic
6'6" (2 m)
0.75" (19 mm)
Heated air tubing temperature cut-out:
≤
106°F (
≤
41°C)
Notes:
•
The manufacturer reserves the right to change these specifications without notice.
•
The electrical connector end of the heated air tubing is only compatible with the air outlet at the device end and should
not be fitted to the mask.
•
Do not use electrically conductive or antistatic air tubing.
•
The temperature and relative humidity settings displayed are not measured values.
Displayed values
Value
Range
Display resolution
Pressure sensor at air outlet:
Mask pressure
4–20 cm H
2
0 (4–20 hPa)
0.1 cm H
2
O (0.1 hPa)
Flow derived values:
Leak
0–120 L/min
1 L/min
Value
Accuracy
Pressure measurement
1
:
Mask pressure
2
±[0.5 cm H
2
0 (0.5 hPa) + 4% of measured value]
Flow and flow derived values
1
:
Flow
±6 L/min or 10% of reading, whichever is greater, at 0 to 150 L/min positive flow
Leak
2
±12 L/min or 20% of reading, whichever is greater, 0 to 60 L/min
1
Results are expressed as STPD (Standard Temperature and Pressure, Dry).
2
Accuracy may be reduced by the presence of leaks, supplemental oxygen, tidal volumes <100 mL or minute ventilation <3 L/min.
3
Measurement accuracy verified as per EN ISO 10651-6:2009 for Home Care Ventilatory Support Devices (Figure 101 and Table 101) using
nominal ResMed mask vent flows.
Measurement system uncertainties
In accordance with ISO 80601-2-70:2015 the measurement uncertainty of the manufacturer's test equipment is:
For measures of flow
± 1.5 L/min or ± 2.7% of reading (whichever is greater)
For measures of volume (< 100 mL)
± 5 mL or 6% of reading (whichever is greater)
For measures of volume (
≥
100 mL)
± 20 mL or 3% of reading (whichever is greater)
For measures of static pressure
± 0.15 cm H
2
O (hPa)
For measures of dynamic pressure
± 0.27 cm H
2
O (hPa)
For measures of time
± 10 ms
Note: ISO 80601-2-70:2015 stated accuracies and test results provided in this manual for these items already include the
relevant measurement uncertainty from the table above.