Water capacity:
To maximum fill line 380 mL
Standard water tub - material:
Injection molded plastic, stainless steel and silicone seal
Cleanable water tub - material:
Injection molded plastic, stainless steel and silicone seal
Temperature
Maximum heater plate:
154°F (68°C)
Cut-out:
165°F (74°C)
Maximum gas temperature:
≤
106°F (
≤
41°C)
Air filter
Standard:
Material: Polyester non woven fiber
Average arrestance: >75% for ~7 micron dust
Hypoallergenic:
Material: Acrylic and polypropylene fibers in a polypropylene
carrier
Efficiency: >98% for ~7-8 micron dust; >80% for ~0.5 micron
dust
Aircraft use
ResMed confirms that device meets the Federal Aviation Administration (FAA) requirements (RTCA/DO-160, section 21,
category M) for all phases of air travel.
Wireless module
Technology used:
3G (USA and Canada only)
2G GSM (all regions except USA and Canada
IC ID: 9103A-AIR103G (USA and Canada only)
The AirCurve 10 device complies with Industry Canada license-exempt RSS standards.
The AirCurve 10 device should be used at a minimum distance of 0.8” (2 cm) from the body during operation.
Additional information regarding IC compliance for this device can be found on www.resmed.com/downloads/devices.
FCC ID: 2ACHL-AIR103G
The AirCurve 10 device complies with FCC Rules.
The AirCurve 10 device should be used at a minimum distance of 0.8” (2 cm) from the body during operation.
Additional information regarding the FCC Rules for this device can be found on www.resmed.com/downloads/devices.
Operating pressure range
ASV, ASVAuto:
4 to 25 cm H
2
O
CPAP
4 to 20 cm H
2
O
Supplemental Oxygen
Maximum flow:
15 L/min (CPAP, ASV); 4 L/min (ASVAuto)
Pneumatic flow path
1. Flow sensor
2. Blower
3. Pressure sensor
4. Mask
5. Air tubing
6. Water tub
7. Device
8. Inlet filter
Design life
Device, power supply unit:
5 years
Cleanable water tub:
2.5 years
Standard water tub, air tubing:
6 months
General
The patient is an intended operator.
English
17
Содержание AirCurve10 ASV