Technical specifications 

90W power supply unit 

AC input range: 

100–240V, 50–60Hz 1.0–1.5A, Class II 

115V, 400Hz 1.5A, Class II (nominal for aircraft use) 

DC output: 

24V 

  3.75A 

Typical power consumption: 

53W (57VA) 

Peak power consumption: 

104W (108VA) 

 
 

Environmental conditions 

 

Operating temperature: 

+41°F to +95°F (+5°C to +35°C) 

 

Note: 

The air flow for breathing produced by this therapy 

device can be higher than the temperature of the room. 

Under extreme ambient temperature conditions 

(104°F/40°C) the device remains safe. 

Operating humidity: 

10 to 95% relative humidity, non-condensing 

Operating altitude: 

Sea level to 8,500' (2,591 m); air pressure range 1013 hPa to 

738 hPa 

Storage and transport temperature: 

-4°F to +140°F (-20°C to +60°C) 

Storage and transport humidity: 

5 to 95% relative humidity, non-condensing 

 
 

Electromagnetic compatibility 

The AirCurve 10 complies with all applicable electromagnetic compatibility requirements (EMC) according to 

IEC60601-1-2:2007, for residential, commercial and light industry environments. It is recommended that mobile 

communication devices are kept at least 1 m away from the device. 
Information regarding the electromagnetic emissions and immunity of this ResMed device can be found on 

www.resmed.com, on the Products page under Service and Support. 

 
 

IEC 60601-1:2005 classification 

Class II (double insulation), Type BF, Ingress protection IP22. 

 
 

Sensors 

Pressure sensor: 

Internally located at device outlet, analog gauge pressure 

type, -5 to +45 cm H

2

Flow sensor: 

Internally located at device inlet, digital mass flow type, -70 

to +180 L/min 

 
 

Maximum single fault steady pressure 

Device will shut down in the presence of a single fault if the steady state pressure exceeds: 
30 cm H

2

O for more than 6 sec or 40 cm H

2

O for more than 1 sec. 

 
 

Sound 

Pressure level measured according to ISO 17510-1:2007 (CPAP mode): 
SlimLine: 

26.6 dBA with uncertainty of 2 dBA 

Standard: 

26.6 dBA with uncertainty of 2 dBA 

Power level measured according to ISO 17510-1:2007 (CPAP mode): 
SlimLine: 

34.6 dBA with uncertainty of 2 dBA 

Standard: 

34.6 dBA with uncertainty of 2 dBA 

Declared dual-number noise emission values in accordance with ISO 4871:1996. 

 
 

Physical - device and water tub 

Dimensions (H x W x D): 

4.57" x 10.04" x 5.91" 

(116 mm x 255 mm x 150 mm)   

Air outlet (complies with ISO 5356-1:2004): 

22 mm 

Weight (device and standard water tub): 

44 oz (1248 g) 

Weight (device and cleanable water tub): 

44 oz (1248 g) 

Housing construction: 

Flame retardant engineering thermoplastic 

16

 

Содержание AirCurve10 ASV

Страница 1: ...User guide English User guide...

Страница 2: ...with chronic symptomatic heart failure NYHA 2 4 with reduced left ventricular ejection fraction LVEF 45 and moderate to severe predominant central sleep apnea Positive airway pressure therapy may be...

Страница 3: ...se with the device including Air tubing heated and non heated ClimateLineAir SlimLine Standard Water tub Standard water tub cleanable water tub can be disinfected Filter Hypoallergenic filter standard...

Страница 4: ...select an option Turn to adjust a selected option and press to save your change Home button Press to return to the Home screen Different icons may be displayed on the screen at different times includi...

Страница 5: ...to the power outlet 3 Connect the air tubing firmly to the air outlet located on the rear of the device 4 Open the water tub and fill it with distilled water up to the maximum water level mark Do not...

Страница 6: ...that adjusts the screen brightness based on the light in the room Stopping therapy 1 Remove your mask 2 Press Start Stop or if SmartStart is enabled therapy will stop automatically after a few seconds...

Страница 7: ...e you can personalize your options Ramp Time Designed to make the beginning of therapy more comfortable Ramp Time is the period during which the pressure increases from a low start pressure to the pre...

Страница 8: ...air 3 Adjust the mask mask cushion and headgear until you get a Good result To stop Mask Fit press the dial or Start Stop If you are unable to get a good mask seal talk to your care provider More opti...

Страница 9: ...ining water 3 Hold the cuff of the air tubing and gently pull it away from the device 4 Hold both the cuff of the air tubing and the swivel of the mask then gently pull apart Cleaning You should clean...

Страница 10: ...water tub and air tubing are dry you can reassemble the parts 1 Connect the air tubing firmly to the air outlet located on the rear of the device 2 Open the water tub and fill it with distilled room t...

Страница 11: ...and send it back to your care provider For more information on the SD card refer to the SD card protective folder provided with your device Note The SD card should not be used for any other purpose Tr...

Страница 12: ...troubleshooting Problem possible cause Solution Air is leaking from around my mask Mask may be fitted incorrectly Make sure your mask is fitted correctly See your mask user guide for fitting instructi...

Страница 13: ...d where there is coverage ie on your bedside table not in a drawer or on the floor The Wireless signal strength icon indicates good coverage when all bars are displayed and poor coverage when fewer ba...

Страница 14: ...e lock read only position Move the switch on the SD Card from the lock position to the unlock position and then re insert it System fault refer to user guide Error 004 Device may have been left in a h...

Страница 15: ...and plug are in good condition and the equipment is not damaged Keep the power cord away from hot surfaces If you notice any unexplained changes in the performance of the device if it is making unusu...

Страница 16: ...aled air Make sure that the mask vent holes are kept clear and unblocked to maintain the flow of the fresh air into the mask Be careful not to place the device where it can be bumped or where someone...

Страница 17: ...he electromagnetic emissions and immunity of this ResMed device can be found on www resmed com on the Products page under Service and Support IEC 60601 1 2005 classification Class II double insulation...

Страница 18: ...3G USA and Canada only The AirCurve 10 device complies with Industry Canada license exempt RSS standards The AirCurve 10 device should be used at a minimum distance of 0 8 2 cm from the body during op...

Страница 19: ...lly conductive or antistatic air tubing The temperature and relative humidity settings displayed are not measured values Displayed values Value Range Display resolution Pressure sensor at air outlet M...

Страница 20: ...0 5 0 8 0 8 12 0 5 0 5 0 5 0 5 0 8 0 8 16 0 5 0 5 0 5 0 5 0 8 0 8 20 0 5 0 5 0 5 0 5 0 8 0 8 25 0 4 0 3 0 6 0 5 0 8 0 8 Symbols The following symbols may appear on the product or packaging Read instr...

Страница 21: ...your ResMed product shall be free from defects in material and workmanship from the date of purchase for the period specified below Product Warranty period Mask systems including mask frame cushion h...

Страница 22: ...incidental or consequential damages claimed to have resulted from the sale installation or use of any ResMed product Some regions or states do not allow the exclusion or limitation of incidental or c...

Страница 23: ...96 Jubilee Ave Milton Park Abingdon Oxfordshire OX14 4RW UK See www resmed com for other ResMed locations worldwide For patent information see www resmed com ip AirCurve Air10 ClimateLine HumidAir and...

Отзывы: