User guide

English

User guide

Содержание AirCurve10 ASV

Страница 1: ...User guide English User guide...

Страница 2: ...with chronic symptomatic heart failure NYHA 2 4 with reduced left ventricular ejection fraction LVEF 45 and moderate to severe predominant central sleep apnea Positive airway pressure therapy may be...

Страница 3: ...se with the device including Air tubing heated and non heated ClimateLineAir SlimLine Standard Water tub Standard water tub cleanable water tub can be disinfected Filter Hypoallergenic filter standard...

Страница 4: ...select an option Turn to adjust a selected option and press to save your change Home button Press to return to the Home screen Different icons may be displayed on the screen at different times includi...

Страница 5: ...to the power outlet 3 Connect the air tubing firmly to the air outlet located on the rear of the device 4 Open the water tub and fill it with distilled water up to the maximum water level mark Do not...

Страница 6: ...that adjusts the screen brightness based on the light in the room Stopping therapy 1 Remove your mask 2 Press Start Stop or if SmartStart is enabled therapy will stop automatically after a few seconds...

Страница 7: ...e you can personalize your options Ramp Time Designed to make the beginning of therapy more comfortable Ramp Time is the period during which the pressure increases from a low start pressure to the pre...

Страница 8: ...air 3 Adjust the mask mask cushion and headgear until you get a Good result To stop Mask Fit press the dial or Start Stop If you are unable to get a good mask seal talk to your care provider More opti...

Страница 9: ...ining water 3 Hold the cuff of the air tubing and gently pull it away from the device 4 Hold both the cuff of the air tubing and the swivel of the mask then gently pull apart Cleaning You should clean...

Страница 10: ...water tub and air tubing are dry you can reassemble the parts 1 Connect the air tubing firmly to the air outlet located on the rear of the device 2 Open the water tub and fill it with distilled room t...

Страница 11: ...and send it back to your care provider For more information on the SD card refer to the SD card protective folder provided with your device Note The SD card should not be used for any other purpose Tr...

Страница 12: ...troubleshooting Problem possible cause Solution Air is leaking from around my mask Mask may be fitted incorrectly Make sure your mask is fitted correctly See your mask user guide for fitting instructi...

Страница 13: ...d where there is coverage ie on your bedside table not in a drawer or on the floor The Wireless signal strength icon indicates good coverage when all bars are displayed and poor coverage when fewer ba...

Страница 14: ...e lock read only position Move the switch on the SD Card from the lock position to the unlock position and then re insert it System fault refer to user guide Error 004 Device may have been left in a h...

Страница 15: ...and plug are in good condition and the equipment is not damaged Keep the power cord away from hot surfaces If you notice any unexplained changes in the performance of the device if it is making unusu...

Страница 16: ...aled air Make sure that the mask vent holes are kept clear and unblocked to maintain the flow of the fresh air into the mask Be careful not to place the device where it can be bumped or where someone...

Страница 17: ...he electromagnetic emissions and immunity of this ResMed device can be found on www resmed com on the Products page under Service and Support IEC 60601 1 2005 classification Class II double insulation...

Страница 18: ...3G USA and Canada only The AirCurve 10 device complies with Industry Canada license exempt RSS standards The AirCurve 10 device should be used at a minimum distance of 0 8 2 cm from the body during op...

Страница 19: ...lly conductive or antistatic air tubing The temperature and relative humidity settings displayed are not measured values Displayed values Value Range Display resolution Pressure sensor at air outlet M...

Страница 20: ...0 5 0 8 0 8 12 0 5 0 5 0 5 0 5 0 8 0 8 16 0 5 0 5 0 5 0 5 0 8 0 8 20 0 5 0 5 0 5 0 5 0 8 0 8 25 0 4 0 3 0 6 0 5 0 8 0 8 Symbols The following symbols may appear on the product or packaging Read instr...

Страница 21: ...your ResMed product shall be free from defects in material and workmanship from the date of purchase for the period specified below Product Warranty period Mask systems including mask frame cushion h...

Страница 22: ...incidental or consequential damages claimed to have resulted from the sale installation or use of any ResMed product Some regions or states do not allow the exclusion or limitation of incidental or c...

Страница 23: ...96 Jubilee Ave Milton Park Abingdon Oxfordshire OX14 4RW UK See www resmed com for other ResMed locations worldwide For patent information see www resmed com ip AirCurve Air10 ClimateLine HumidAir and...