Troubleshooting
Problem/possible cause
Solution
Mask is uncomfortable
Headgear straps are too tight.
The cushion membrane is designed to
inflate against the patient's face to provide
a comfortable seal with low headgear
tension. Adjust straps evenly. Ensure that
the headgear straps are not too tight and
that the cushion is not crushed or creased.
Mask could be the wrong size.
Check the patient's size against the fitting
template located on the mask packaging.
Mask leaks around the face.
Cushion membrane is creased or folded.
Refit the mask according to the instructions.
Ensure that you position the cushion
correctly on the patient's face before pulling
headgear over their head. Do not slide the
mask down the patient's face during fitting
as this may lead to folding or curling of the
cushion.
Mask could be the wrong size.
Check the patient's size against the fitting
template located on the mask packaging.