KrosFlo® TFF System – Auxiliary Scale ACSS-60K
User Guide
18 repligen.com 420-14871-002 Rev. 00
7.1.3
Printout interval for continuous transmission
Parameter to set frequency of printout for continuous transmission. Printout interval is set in
seconds, with 0.1 [s] accuracy within 0.1 [s] – 3600 [s] range.
1.
Enter
<P5.1.PC / 5.1.3.Int>
submenu.
2.
Enter transmission rate, to do this press
to select digit to be edited and
to set
digit value (0 - 9).
3.
Set respective value and press
key to confirm, next go to the home screen.
7.1.4
Printer
<5.2.Prtr>
submenu allows to:
•
Select port to which the printer is connected
7.1.5
Port
1.
Enter
<P5.2.Prtr / 5.2.1.Prt>
submenu.
2.
Press
key, options are displayed successively, where:
nonE
– No port selected,
rS1
–
RS232.
3.
Set respective value and press
key to confirm, next go to the home screen.
7.2
Additional display
<5.3.AdSP>
submenu allows to:
•
Select port to which the additional display is connected
7.2.1
Port
1.
Enter
<P5.3.AdSP / 5.3.1.Prt>
submenu.
2.
Press
key, options are displayed successively, where:
nonE
– No port selected,
rS1
–
RS232 (1).
3.
Set respective value and press
key to confirm, next go to the home screen.
8.
Printouts
8.1
Calibration report
<P6.1.CrEP>
is a group of parameters allowing to declare variables that are to be printed on the
calibration report printout. Each variable allows for accessibility attribute:
YES
– print,
no
– do not
print. Calibration report is automatically generated at the end of each calibration process.