Practice Mode and Training Mode Measurements
Measure
Range
Accuracy
Precision
Angle
0-180°
+/- 0. 2°
1°
Start Angle
0-180°
+/- 0. 2°
1°
Max Angle
0-180°
+/- 0.2°
1°
Session Score
0-100
+/- 0.5
1
Final Score
0-100
+/- 0.5
1
Storing Your leva
Storage of the leva PDHS should not exceed the following conditions:
•
Temperature: +5°C to +40°C
•
Humidity: 90% RH non-condensing
•
Pressure/Altitude: 3000m or 70kPa
Operating Your leva
Operation of the leva PDHS should not exceed the following
conditions:
•
Temperature: +5°C to +40°C
•
Humidity: 90% RH non-condensing
•
Pressure: 700hPa to 1060hPa
•
Operating Range: <3.2 ft (1 meter)
Expected Service Life
•
The expected service life of the leva PDHS is 1 year.
•
The leva PDHS requires periodic battery changes to maintain
reliability (see Replacing Batteries in the Maintenance Section).
•
App updates will be pushed to users periodically as needed
through the App Store™ or Google Play™.
Maintenance
Use the following instructions to ensure proper care and maintenance of
your leva PDHS.
Cleaning the Probe:
To ensure hygienic safety, you should clean and dry your leva probe before
and after each treatment session.
•
Clean the leva probe with mild, warm, soapy water.