Renke RS-MG41-1 Скачать руководство пользователя страница 12

RS-MG41-1 Composite Multi-Gas Detector User Manual V1.0

Shandong Renke Control Technology Co., Ltd.

www.renkeer.com

12

5.3.3 Store data

Device View

Note: The storage settings need to be turned on in the system settings. The device cannot view the
specific historical data. You need to use the configuration soft armor to read the data. Can be
viewed in the device: gas type, current value, range, maximum value, minimum value, start
storage time, end Storage time and number of storages.

Содержание RS-MG41-1

Страница 1: ...RS MG41 1 Composite Multi Gas Detector User Manual V1 0 Shandong Renke Control Technology Co Ltd www renkeer com 1 RS MG41 1 Composite multi gas detector User manual Document version V1 0...

Страница 2: ...rrying 6 5 Function and operation instructions 7 5 1Key Description 7 5 2Main interface introduction 8 5 3Instructions 9 5 3 1System setup instructions 9 5 3 2Channel setting instructions 11 5 3 3Stor...

Страница 3: ...rial site safety monitoring for high reliability of the equipment The housing is made of high strength engineering plastics and composite non slip rubber which has high strength and good feel and is w...

Страница 4: ...y safe explosion proof products 3 High definition color screen design easy to check the gas status 4 The product can detect four gases oxygen flammable gas carbon monoxide and hydrogen sulfide 5 Can s...

Страница 5: ...OL Indication error Absolute error 5ppm 5 ppm 5 FS 3 FS Relative error 10 Duplicate value 2 2 2 1 Zero drift 3ppm 5ppm 2 FS 1 Span drift 5 5ppm 3 FS 1 Alarm action value 50 150ppm 10 35ppm 20 50 LEL 1...

Страница 6: ...pment USB plug USB data cable standard gas hood certification warranty card factory inspection report 4 Way of carrying Open the device with the surface buckle snap the device to the place where you n...

Страница 7: ...face short press to increase current value long press to increase current value quickly Decrease key Main interface long press to turn off the buzzer alarm Menu bar interface short press the cursor mo...

Страница 8: ...current option 5 2Main interface introduction Serial number name Description 1 Real time value Display the current actual gas concentration value 2 Gas name Show gas name 3 time Display the current ti...

Страница 9: ...annot be enabled 11 Proportion Display the current remaining battery level 5 3Instructions 1 In the shutdown state press and hold the 2S confirmation key the device detects whether the buzzer flash an...

Страница 10: ...m doesn t affect the low battery alarm function Standby screen Select the corresponding time click the OK button Password setup With increase button and reduce button to confirm the present value the...

Страница 11: ...ore than H Alarm value it will has has Alarm II Zero standard concentration Note this operation must be operated in the clear air or N2 Enter in the operation screen please move the cursor to Confirm...

Страница 12: ...ice View Note The storage settings need to be turned on in the system settings The device cannot view the specific historical data You need to use the configuration soft armor to read the data Can be...

Страница 13: ...parameters Connect the device to the computer through the USB data cable click Read Device Information in the toolbar and click Write Device Information after editing the device parameters to write th...

Страница 14: ...hen an alarm occurs on the detector If you do not allow it you cannot turn off the alarm sound quickly Sound on Set the alarm sound of the detector on or off Alarm on Set the detector s sound light an...

Страница 15: ...nd then connect the recorder to the computer via a USB cable Click Read Data to import the data stored in the tester into the configuration software Click on the data to view the stored historical dat...

Страница 16: ...k Delete device storage data in the drop down menu to clear the device data After clearing the software prompts Delete cleared successfully 7 Gas calibration Due to the normal loss and aging of the se...

Страница 17: ...be used normally 9 Precautions 1 Prevent the machine from falling from high places or receiving severe vibration 2 In high concentration gas environment the machine may not work properly 3 Please use...

Страница 18: ...m detection gas circuit failure Contact your dealer or manufacturer for repairs Inaccurate display Sensor expired Contact your dealer or manufacturer Long uncalibrated Replace the sensor Time display...

Страница 19: ...ntact information Shandong Renke Measurement Control Technology Co Ltd Address 2 F East Block Building 8 Shun Tai Plaza High tech Zone Jinan City Shandong Province Postcode 250101 Phone 400 085 5807 F...

Страница 20: ...ppm Low 5 High 10 ppm SO2 0 20ppm 0 50 1000ppm 0 1 1ppm Low 5 High 10 ppm CL2 0 20ppm 0 100 1000ppm 0 1ppm Low 5 High 10 ppm Hydrogen H2 0 1000ppm 0 5000ppm 1ppm Low 50 High 150 ppm Ammonia 0 100ppm 0...

Отзывы: