TONiC RKLC20-S linear installation guide
4
TONiC interface drawing
CAL button operation
Push and release (<3 seconds) - Calibration (CAL) routine enable/disable
Push and release (>3 seconds) - Automatic Gain Control (AGC) enable/disable
Push and hold during power ‘Off / On’ cycle - Restore factory defaults
Refer to readhead LED functionality chart for CAL LED indications
15
1
8
9
16
4-40 UNC
× 2
40
6 min
62
67
CAL /AGC push button switch
access hole Ø2.4
33.3
R>20 Dynamic bend radius*
R>10 Static bend radius
8
Diagnostic LED
(Ti0004 - Ti20KD
and
TD4000 - TD0040)
Cover plate
*
Dynamic bend radius not applicable for UHV cables.
Dimensions and tolerances in mm