4.6
Trig
g
er Logic
RMP600 – RMI partnership
System set-up is achieved using Trigger Logic and powering the RMI. Partnering is only required
during initial system set-up. Further partnering will be required if either the RMP600 or RMI is
changed, or if a system is reconfigured for multiple probes (
multiple probe mode
).
Partnering will not be lost by reconfiguring the probe settings or when changing batteries, except
where
“multiple probe mode”
is selected. Partnering can take place anywhere within the operating
envelope.
In configuration mode, configure the probe settings as required until you reach the
“
Acquisition
mode”
menu, which defaults to
“
Acquisition
mode off”
.
Deflect the stylus to select
“
Acquisition
mode on”
. Ensure this is done within
8 seconds of the RMI signal LED
flashing green.
The probe is now in standby
and the system is ready for use.
> 20 s
SIGNAL LED
RMI in
“Acquisition mode”
SIGNAL LED
New partner radio probe
acquired
Acquisition mode
Acquisition mode off
Switch the RMI on
< 4 s
NOTE:
Refer to the
RMI radio
machine interface
installation
guide (Renishaw part no.
H-4113-8554) when partnering