Fault finding
Table - False (‘air’) triggers occur at traverse speed and the probe LEDs
flicker.
PI 200 indicators:
‘DAMPED’ LED is ON.
‘SEATED’ LED operates normally.
Possible cause
Checks/remedies
• The stylus is too large or too
heavy.
• There is excessive vibration from
the CMM.
• The traverse speed is too high.
• Use stylus arrangements within
recommendations.
• Check the CMM’s air supply.
• Maintain the CMM’s air bearing system.
• Reduce the traverse speed.
Table 7 - The probe triggers during an SCR200 stylus change.
PI 200 indicators:
‘SEATED’ LED operates normally.
Possible cause
Checks/remedies
• The SCR200 is not connected to
the PI 200.
• Incorrect SCR200 operating
mode.
• Check the SCR200 indicator lamps.
• Reconnect the cable.