12
QUANTiC RTLC40 /
FASTRACK
installation guide
Readhead mounting and alignment
Mounting brackets
The bracket must have a flat mounting surface and should provide adjustment to enable conformance to the
installation tolerances, allow adjustment to the rideheight of the readhead, and be sufficiently stiff to prevent
deflection or vibration of the readhead during operation.
Readhead set-up
Ensure that the scale, readhead optical window and mounting face are clean and free from obstructions.
NOTE:
When cleaning readhead and scale apply cleaning fluid sparingly, do not soak.
To set nominal rideheight, place the spacer with the aperture under the optical centre of the readhead to allow
normal LED function during set-up procedure. Adjust the readhead to achieve a flashing Green LED along
the full axis of travel. The faster the flash rate, the closer it is to optimum set-up. The optional Advanced
Diagnostic Tool ADTi-100 (A-6195-0100) and ADT View software can be used to optimise signal strength in
challenging installations. See
for more information.
NOTE:
When re-installing the readhead factory defaults should be restored
.
Green Orange Red
flashing flashing flashing
Readhead set-up LED status
Roll
0° ±0.8°
Yaw
0° ±0.9°
Pitch
0° ±1°
Rideheight 1.9 ±0.17 mm
from
FASTRACK
top surface
Black spacer
Readhead LED diagnostics*
Mode
LED
Status
Installation
mode
Green flashing
Good set-up, maximise flash rate for optimum set-up
Orange flashing
Poor set-up, adjust readhead to obtain Green
flashing LED
Red flashing
Poor set-up, adjust readhead to obtain Green
flashing LED
Calibration
mode
Blue single flashing
Calibrating incremental signals
Blue double flashing
Calibrating reference mark
Normal
operation
Blue
AGC on, optimum set-up
Green
AGC off, optimum set-up
Red
Poor set-up; signal may be too low for reliable
operation
Blank flash
Reference mark detected
(visible indication at speed < 100 mm/s only)
Alarm
Four red flashes
Low signal, over signal, or overspeed;
system in error
Red and Purple
flashing (analogue
variant only)
AGC out of normal operating range
*
for more information on diagnosing faults.