13
Technical data
4.
Technical data
4.1
Gauge Checker test limits
The test limits set for the Gauge Checker have been derived
from the performance specifications quoted for the Equator and
the measurement uncertainty
†
evaluated for the inspection of the
Gauge Checker (to a 95% confidence interval).
•
Size
1
: ± 2 µm
•
Position
1
: ± 2 µm
•
Distance:
•
1D
2
: ± 2.8 µm
•
2D
3
: ± 4.0 µm
•
3D
4
: ± 4.9 µm
•
Form error
5
: 5 µm
•
Freeform curve deviation
6
: ± 5 µm
Note:
These limits have been set for the Golden Part
Comparison method, with no calibration file (.cal file), and with
an assumption of no more than 2 °C temperature difference
between master and measure.
1
Size and position: the Equator performance specification
quoted is ±2 µm (evaluated using empirical measurements)
2
1D Distance:
√
2
2
+ 2
2
= 2
√
2 = ± 2.828 µm