12
•
Open the Gauge Checker wizard.
•
Follow the steps within the Gauge Checker wizard. When
the Gauge Checker process has finished, a pass/fail report
will be produced.
3.3
Interpreting the results
The pass/fail report shows deviations from nominal for each
feature during the measure cycle and these deviations are tested
against the comparison uncertainty limits for each feature.
Pass
- If the deviation is within the comparison uncertainty limits
the result is a pass.
Fail
- When the results are outside the comparison uncertainty
the result is a fail. If an Equator fails, it is most likely to be down
to dirt on the equipment. Clean the Gauge Checker artefact
and stylus and run the test again. If the test fails a second time,
contact your local Renishaw distributor.
Multiple measure cycles
- When performing multiple measure
cycles, it is possible that thermal effects could give rise to larger
deviations.
Pass/fail will therefore only be tested against the first measure
cycle, but the use of multiple measure cycles can be a good way
to assess the stability of your inspection process in a changing
thermal environment.
Product operation
Содержание Equator Gauge Checker
Страница 1: ...User s guide Equator Gauge Checker H 5925 8600 01 B Equator Gauge Checker ...
Страница 3: ...1 Equator Gauge Checker Installation and user s guide ...
Страница 6: ...4 This page intentionally left blank ...
Страница 12: ...10 C D E E A F Product operation ...
Страница 17: ...15 Technical data ...
Страница 18: ...16 Technical data ...