- 2 -
Error
Cause
Solution
No vacuum being gen
-
erated.
•
Bowl detection pin sticking.
• Clean area around pin.
• Have the device repaired.
Permanent venting
noise.
• Solenoid valve fault.
• Have the device repaired.
•
Bowl detection pin sticking.
• Clean area around pin.
• Have the device repaired.
Fluttering noise from
the silencer.
•
Working pressure too low.
•
Check dynamic operating pressure
(see technical data, Sec. 9).
Error codes
Error code
Cause
Solution
Err. 2
•
Minimum vacuum level not yet
reached.
•
Vacuum level has dropped be
-
low 500 mbar.
•
Ventilate bowl and remove (see section
5.4.1, points no. 2 and 3).
•
If the error recurs:
Have the device repaired.
Err. 3
•
The vacuum is built-up too
quickly due to the fact that the
suction
system is blocked.
•
Have filter cleaned or changed (Sec.
7.3).
Err. 4
• Motor fault.
•
Ventilate bowl and remove (see Sec.
5.4.1, points no. 2 and 3).
• Have the device repaired.
Err. 10
•
Too much material.
•
Mixing time too long, the materi
-
al has already begun to set.
•
Ventilate bowl and remove (see section
5.4.1, points no. 2 and 3).
•
Only fill the bowl to the maximum fill
level.
This limit is valid for powders and liquids
measured before mixing.
•
Ventilate bowl and remove (see section
5.4.1, points no. 2 and 3).
•
Select shorter mixing times.