© R.E.Morrison Equipment Inc. 2012

BaseVac™ M-Vac Series Installation Guide

|

2

Important 

Information

Intended Use

To provide vacuum suction during general examinations and procedures conducted by qualified dental professionals.

Disposal of Equipment

At the end of product life, the unit(s), accessories, and other consumable goods may become contaminated from normal 
use. Consult local codes and ordinances for proper disposal of equipment, accessories and other consumable goods.

Transportation / Storage Conditions

Ambient Temperature Range...............................0°F to 160°F (-17°C to +71°C)
Relative Humidity ................................................10% to 90% (non-condensing)
Atmospheric Pressure .........................................500hPa to 1060hPa (0.49atm to 1.05atm)

Ventilation Information

Dry rotary vanes pumps run hot and can generate heat inside a small room, to reduce room internal temperature an exhaust 
fan may be needed, see below for fan size details: The general rules are 2500 BTU per Hp or about 230 CFM of ventilation 
per Hp to keep the room at 10° F above outside ambient.

Safety Symbols

Danger

Indicates an imminently hazardous situation which will 
result in serious or fatal injury if not avoided. This symbol is 
used only in the most extreme conditions.

Warning

Indicates a potentially hazardous situation which could 
result in serious injury if not avoided.

Caution

Indicates a potentially hazardous situation which may 
result in minor or moderate injury if not avoided. It may also 
be used to alert against unsafe practices.

Equipment Alert

Indicates a potentially hazardous situation which could 
result in equipment damage if not avoided.

Note

Amplifies a procedure, practice, or condition.

REM Equipment Inc.
3615 Laird Road

Mississauga, ON  L5L 5Z8 

Canada

T 1.800.668.8736 
F 905.828.3674 
[email protected] 
www.remequip.com

BaseVac™ Systems are Manufactured by:

Содержание BaseVac 1M4.10

Страница 1: ... 1 800 668 8736 www basevacdental com Manufactured by R E Morrison Equipment Inc an ISO 9001 and ISO 13485 certified company M Vac Series For models 1M4 10 1M4 16 1M4 25 2M4 10 2M4 16 2M4 25 Installation Guide ...

Страница 2: ...temperature an exhaust fan may be needed see below for fan size details The general rules are 2500 BTU per Hp or about 230 CFM of ventilation per Hp to keep the room at 10 F above outside ambient Safety Symbols Danger Indicates an imminently hazardous situation which will result in serious or fatal injury if not avoided This symbol is used only in the most extreme conditions Warning Indicates a po...

Страница 3: ...e Level DBA 3 64 64 67 67 67 67 Dimensions L 17 x W 17 x H 30 L 17 x W 17 x H 30 L 17 x W 17 x H 30 L 17 x W 17 x H 30 L 17 x W 17 x H 39 L 17 x W 17 x H 39 Weight lbs 4 72 72 86 86 98 98 Air Water Separator Built In Built In Built In Built In Built In Built In INLET 1 50 Hose Barb DRAIN OUTLET 1 50 Plastic Pipe Only 2M models come complete with remote start stop capabilities 24 volt Electronic hi...

Страница 4: ... R E Morrison Equipment Inc 2012 BaseVac M Vac Series Installation Guide 4 Loose Parts x1 x2 x1 x1 x1 x1 5 x2 4 x1 10 23 x1 x1 x1 x1 Included with 220 single phase versions ...

Страница 5: ... Connect the tank drain outlet hardware to the drain connection located at the bottom of the tank Make sure that the gravity check valve flap is hanging down The check valve has a 1 5 PVC compression fitting for connection to a drain drain piping is not supplied These fittings should be fastened with PVC glue Note 1 1 Tank must be elevated above drain All connection hardware except for drain pipin...

Страница 6: ...st piping should run to an outdoor source Assemble piping as shown Use PVC glue to fasten piping Extension to outside is to be supplied by others 1 2 Connect hose barb located on bottom of p trap and elbow to drain Hose is not supplied DropLeg with barded fitting to drain hose Supplied by Others To outside source ...

Страница 7: ...d For a 115 volt connection a minimum of 108 V 1 ph and a maximum of 122 V 1 ph is to be connected If below or above a buck boost transformer should be installed to maintain proper voltage This input voltage should be checked at least 3 times daily to ensure stable voltage in some cases voltages may vary from the load in the area during different times of day Plug receptacle into 15 amp protected ...

Страница 8: ... R E Morrison Equipment Inc 2012 BaseVac M Vac Series Installation Guide 8 Switch Wiring for Version 1 Systems ...

Страница 9: ...BaseVac M Vac Series Installation Guide 9 www basevacdental com Version 2 optional Electrical ...

Страница 10: ...ing vane replacement guide Insure vacuum regulating valve is turned all the way in Pump room is excessively hot Exhaust piping The pump room heats up during the day Look for and correct leaks crack or separation in exhaust or piping Remove outside obstructions example bird nest garbage receptacle Liquid leaking from front of pump Air water separator malfunction Pump continues to run water leaks fr...

Страница 11: ... ...

Страница 12: ...ntal com Manufactured by R E Morrison Equipment Inc an ISO 9001 and ISO 13485 certified company Our products perform extremely well Our company performs even better We work to ensure you are satisfied with the performance of our products today everyday We are here to help ...

Отзывы: