Troubleshooting
38 / 44
TARAtec 9/10
8.2.4
Signal
The zero point checking must have been performed successfully.
Add the relevant disinfectant to the mains water in the glass beaker
(see section 8.2.3, p. 37).
Stir the sensor steadily round in the glass beaker for five minutes.
Monitor the measured value throughout this time.
If the measured value increases, the sensor can provisionally be
regarded as OK.
If the measured value does not change:
Perform maintenance on the sensor (see section 7, p. 25) and repeat
the signal test.
This completes the signal test. The sensor can be put back into use.
If after maintenance the sensor shows no response to the disinfectant:
Return the sensor to the supplier for checking.