Troubleshooting
TARAtec 9/10
29 / 44
8.1
Fault overview
Fault
Cause
Corrective action
Sensor cannot
be calibrated /
the measured
value deviates
from the
analytical
measurement
Run-in time too short.
Wait until the run-in
time has elapsed (see
section 4.5, p. 22).
Repeat the calibration
after a few hours.
Membrane torn.
Change the
membrane cap (see
section 7.2, p. 26).
Membrane cap
damaged.
Change the
membrane cap (see
section 7.2, p. 26).
Disruptive substances in
the water contents
Check the water for
disruptive substances
(observe data sheet).
Provide remedy.
Consult the supplier.
Short circuit / defect in
the measuring lead
Locate and eliminate
the short circuit /
defect.
Exchange the
measuring lead.
Distance between
working electrode and
membrane is too great.
Screw the membrane
cap on fully to the
stop.
The DPD/titration
chemicals are out of
date.
Use fresh
DPD/titration
chemicals.
Repeat the calibration
(see section 5, p. 23).