Calibration
TARAtec 10.1
23 / 40
Measured variables
Recommended analytical methods
Chlorine dioxide
DPD-1
Photometer for chlorine
dioxide
Ozone
DPD-1 + DPD-3
Photometer for ozone
DPD-4
Tab. 4: Recommended analytical methods
Chlorine dioxide can also be determined using a photometer
intended for chlorine. The result must be multiplied by a factor of
1.9.
Ozone can also be determined using a photometer intended for
chlorine. The result must be multiplied by a factor of 0.68.
At higher concentrations of disinfectant the DPD colouration may fail
to appear.