Commissioning
TARAtec 10.1
21 / 40
4.4.3
Connection with Modbus signal transmission
The sensor is provided with a 5-pin M12 screwed plug protected against
polarity reversal. There are no termination resistors within the sensor.
The connection pins are assigned as follows:
Fig. 9:
Connection pin assignment (5-pin)
1
(reserved)
2
+9…+30 V
3
Power GND
4
RS485 B
5
RS485 A
4.5
Initial calibration
The sensor must have been connected electrically (see section 4.4,
p. 19).
The run-in time has been complied with as per the data sheet (see
section 1.2, p. 5).
Perform calibration (see section 5, p. 22).
After one day, repeat the calibration.