background image

12      

15140-101 Rev. C

Instructions for Use

Operation 

1.  Turn on the power using the On/Off switch located 

on the front of the Power Supply Assembly.  

Brightness can be adjusted by rotating the 

Illumination Level Control Knob.   

Refer to Figure SU-14.

Note:

   The maximum intensity setting is for 

intermittent use only.  Continuous use will 

shorten the lamp life. 

2.  Insert the Focusing Rod in the pivot post of the 

instrument body to make rough PD and focus 

adjustments.  Refer to Figure IN-1.

3.  Position the light onto the flat surface of the 

Focusing Rod and adjust the pupillary distance 

and focus of the eyepieces to suit the needs of the 

operator.

4.  Using the Slit Width Knobs, adjust the projected 

slit so that the thinnest slit is shown on the 

Focusing Rod.  Refer to Figure IN-2.

Note:

  The thinnest slit will allow for greater accuracy.

5.  Remove the Focusing Rod.

6.  To position a patient, adjust the chinrest height 

by turning the Chinrest Elevation Handle on the 

post of the Chin Rest Assembly until the patient’s 

canthus is in line with the Canthus Mark on the 

chin rest post.  Refer to Figures IN-2 and IN-3.

7.  Microscope elevation is adjusted by rotating the 

joystick and observing the slit image through the 

Microscope Assembly until the slit is centered on 

the patient's cornea.  Refer to Figure IN-4.

8.  Move the slit lamp with the joystick held firmly 

and slightly angled toward the patient, until the slit 

appears sharply on the cornea. 

Note:

  The accuracy of this rough adjustment should 

be checked by the naked eye.  The fine 

adjustment is performed while observing the 

slit through the microscope.

9.  Tilt the joystick, which is now held lightly at its 

upper end, until the slit appears sharply at the 

depth of the eye which is to be observed. 

10. 

The horizontal motion of the base can be locked 

by tightening the base locking screw.  Refer to 

Figure IN-4.

Note:  

Lock the base whenever the lamp is not in use.

Figure IN-1   Install Focusing Rod

Focusing 

Rod

Figure IN-2 Slit Width and Patient Height

Chinrest 

Elevation 

Handle

Slit 

Width 

Knobs

Figure IN-3   Adjust Patient Height

Canthus 

Mark

Figure IN-4   Adjust Height

Adjust Lamp 

Height

Lock 

Base

Содержание XCEL 455

Страница 1: ...XCEL 455 Slit Lamp User s Guide...

Страница 2: ...nges in the product described in this manual without notice and without incorporating those changes in any products already sold ISO 9001 13485 Certified Reichert products are designed and manufacture...

Страница 3: ...3 Filters 13 Slit Rotation 14 Illumination Inclination 14 Slit Centration 14 Cleaning Maintenance 15 Cleaning 15 External Cleaning 15 Forehead Chinrest Cleaning Disinfection 15 Cleaning the Glide Plat...

Страница 4: ...ELECTRIC SHOCK THIS EQUIPMENT MUST ONLY BE CONNECTED TO A SUPPLY MAINS WITH PROTECTIVE EARTH OR DAMAGE TO THIS INSTRUMENT AND OR INJURY TO THE OPERATOR OR PATIENT MAY OCCUR WARNING ENSURE THAT THE VOL...

Страница 5: ...NG EMC AND NEEDS TO BE INSTALLED AND PUT INTO SERVICE ACCORDING TO THE EMC INFORMATION PROVIDED IN THE ACCOMPANYING DOCUMENTS CAUTION PORTABLE AND MOBILE RF COMMUNICATIONS EQUIPMENT CAN AFFECT MEDICAL...

Страница 6: ...014 Date of Manufacture REF Catalog Number S N Serial Number Waste of Electrical and Electronic Equipment Compliance to Medical Device Directive 93 42 EEC 36FS Underwriter s Laboratories has independe...

Страница 7: ...of the anterior segment of the eye Contraindications None Phototoxicity Information Because prolonged intense light exposure can damage the retina the use of the device for ocular examination should n...

Страница 8: ...ial 13 Lamp Housing Cover 14 Fixation Light 15 Filter Lever 16 Slit Rotation Length Knob 17 Slit Rotation Scale 18 Microscope Lock Knob 19 Illumination Arm Lock Knob 20 Instrument Base Lock Knob 21 Gu...

Страница 9: ...he accessories and store them in an appropriate place so that when they are needed they will be available Refer to Figure SU 4 8 Using the 3mm Hex Wrench connect the Ground Wire from the Power Supply...

Страница 10: ...lumination Assembly and Arm Note The Illumination Assembly and Arm are con nected as one piece 14 Mount the Illumination Assembly and Arm onto the Base Assembly and secure it with the Screw using the...

Страница 11: ...NGER TO THE PATIENT WARNING POSITION THIS INSTRUMENT SO THAT IT IS NOT DIFFICULT TO OPERATE THE DISCONNECTION DEVICE PLUG 1 After the unit is in its secure location apply the correct input voltage to...

Страница 12: ...Chin Rest Assembly until the patient s canthus is in line with the Canthus Mark on the chin rest post Refer to Figures IN 2 and IN 3 7 Microscope elevation is adjusted by rotating the joystick and obs...

Страница 13: ...slit length is adjusted by rotating the Slit Length Rotation Dial The dial has five stops for adjustments They are 12 9 5 3 1 and 0 3 mm diameter and continuous length Refer to Figure IN 8 Filters Th...

Страница 14: ...mination Assembly by depressing the Inclination Latch Release and pulling the base of the Illumination Assembly toward the operator Refer to Figure IN 10 Slit Centration When the centering screw is lo...

Страница 15: ...d rest with an alcohol wipe and change the chin rest papers after each patient Cleaning the Glide Plate If the Glide Plate is dirty it may cause a rough feeling when maneuvering the base of the slit l...

Страница 16: ...Holder and Lamp Refer to Figure MM 5 5 Grip the Lamp by the Metal Disk that is attached to it and pull it out of the Lamp Holder Refer to Figure MM 6 6 Replace the Lamp with a new one and install it...

Страница 17: ...gripping the two small tabs Refer to Figures MM 8 and MM 9 3 Open the Door to the Fuse Holder by pulling it down Refer to Figure MM 9 Note The Fuses will pop up when the door is open making removal e...

Страница 18: ...ation Light Harness not plugged into the Power Supply Assembly Ensure the Fixation Light Harness is properly seated in the Power Supply Assembly Defective Power Supply Replace the Power Supply Assembl...

Страница 19: ...35 C 95 F Relative Humidity 30 to 75 Atmospheric Pressure 80 kPa 23 6 in Hg to 106 kPa 31 3 in Hg Transportation Storage Temperature 20 C 4 F to 70 C 158 F Relative Humidity 10 to 80 Atmospheric Pres...

Страница 20: ...tion Protection Class I Ingress Protection Rating IXP0 Instrument Type IEC 60601 1 B Operating Mode Continuous US FDA Class II 21 CFR 886 1850 EU Medical Device Directive Class I Rule 12 CAN CMDR Clas...

Страница 21: ...n such an environment Emissions Test Compliance Electromagnetic Environment Guidance RF Emissions CISPR 11 Group 1 Class A The Xcel 455 uses RF energy only for its internal function Therefore its RF e...

Страница 22: ...ity should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1kV Differential 2kV Common 1kV Differential 2kV Common Mains power quality should be that of a typical commercia...

Страница 23: ...distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range Interference may occur...

Страница 24: ...m output power of the communications equipment Max Output Pow er of Transmitter W Separation m 150kHz to 80 MHz d 3 5 V1 Sqrt P Separation m 80 to 800 MHz d 3 5 E1 Sqrt P Separation m 800MHz to 2 5GHz...

Страница 25: ...RRANTY If notified promptly in writing of any action brought against the purchaser based on a claim that the instrument infringes a U S Patent Reichert will defend such action at its expense and will...

Страница 26: ...26 15140 101 Rev C Notes...

Страница 27: ...27 15140 101 Rev C Notes...

Страница 28: ...716 686 4545 Email reichert information ametek com www reichert com AMETEK GmbH Business Unit Reichert Carl von Linde Strasse 42 85716 Unterschleissheim Munich Germany Email info reichert de ametek c...

Отзывы: