Setup
Attaching the Phoroptor to the Stand
WARNING: PRIOR TO INSTALLING THE PHOROPTOR ONTO THE STAND ARM, VERIFY THAT THE ROD ON
THE STAND ARM IS SECURE BEFORE ATTEMPTING TO INSTALL THE PHOROPTOR OR DAMAGE TO THE
UNIT AND/OR PATIENT MAY OCCUR.
WARNING: REMOVAL OF THE SAFETY SCREW MAY CAUSE THE INSTRUMENT TO DETACH FROM ITS SUP-
PORTING ARM AND CAUSE SERIOUS INJURY.
1. Place the Phoroptor on the instrument stand arm by sliding the Mounting Bracket over the end of
the arm until the threaded hole of the stand arm lines up with the slotted hole in the bottom of the
Mounting Bracket. Refer to Figure SU-1.
2. A Retaining Screw has been provided to prevent the Phoroptor from falling off the stand. Insert
the screw up through the slotted hole and thread it into the stand arm. Tighten the screw firmly in
place. Refer to Figure SU-1.
Note:
The Retaining Screw (P/N 11330-317)
provided with the Phoroptor will work
with most stands. If using an Advan-
tage Stand from Reichert, use Retain-
ing Screw (P/N 15072-032) that comes
with the stand on the Phoroptor Arm.
Note:
The instrument cannot slide off the arm
at this point, but can be tilted forward or
back.
3. Tighten the Tilt Clamp Knob and the
Phoroptor will be held firmly in the desired
position. Refer to Figure SU-1.
Connecting the Power (For Illuminated Phoroptors Only)
1. Insert the male connector into the Power Input receptacle on the Phoroptor Mounting Bracket.
WARNING: CARE MUST BE TAKEN TO ARRANGE THE CABLE FOR THE ADAPTOR SO THAT IT DOES NOT
PRESENT A TRIPPING HAZARD TO THE EXAMINER OR A DANGER TO THE PATIENT.
WARNING: POSITION THIS INSTRUMENT SO THAT IT IS NOT DIFFICULT TO OPERATE THE DISCONNECTION
DEVICE (PLUG).
2. Plug the adapter into an outlet with the correct input voltage.
3. Light should automatically turn on when plugged in.
4. To adjust the brightness of the lights, turn the Rheostat Control Knob. Refer to Page 10.
• Counterclockwise rotation dims the LEDs.
• Clockwise rotation brightens the LEDs.
Note:
The Rheostat Control Knob does not go “OFF,” it only dims and brightens the LED’s.
To shut the lights off entirely you must unplug the adaptor.
Instrument Setup
(continued)
Mounting
Bracket
Tilt Clamp Knob
Stand Arm
Retaining Screw
Figure SU-1 Attaching Phoroptor to Stand