Mary Lift User Manual
03-2015
TD 5.6-001 PR 3500 200 E - ENG ver.4
Side 20 af 20
Annex F Troubleshooting
Nr. Fault
Possible cause
Solution
1
Function failure
Is is switched on?
Switch on
2
Is the battery fully charged?
If no – charge the battery
If yes – check all connections
3
Are all plugs properly connected?
Press hard to connect plug
4
Is the emergency stop activated?
(pressed in)
De-activate the stop button.
Emergency stop button can be
released after use by twisting it a
quarter-turn. The lift is then ready for
use again.
5
Are there visible breaks in/damage to
the cable?
Replace cable
6
If battery cannot be charged
Battery can be defective – contact the
dealer.
Contact the dealer if the equipment is out of order. Do not use defective equipment..