Manual
– TopCon TC.LAE
Product description
11 / 56
2016-06-08
1.3.2.
Overview of the rear side of the device
1
2
3
4
5
7
6
Fig. 5
Overview of interfaces - rear side.
Overview of interfaces - rear side
1
LC inlet ,
Interface nipple G
1/2”
Non drip quick release coupling nipple to connect to the outlet of a device,
which will be cooled by the TC.LAE device.
For further information about the connection procedure,
refer to chapter 6.2.3, page 30.
2
LC outlet,
Interface coupling
G1/2”
Non drop quick release coupling to connect to the inlet of a device,
which will be cooled by the TC.LAE device.
For further information about the connection procedure,
refer to chapter 6.2.3, page 30.
3
LC filling,
interface
G1/2” for the filling hose
Connect the filling hose to this 1/2 ´´thread or to the equal threat on the top
side of the filling block.
For further information, refer to chapter 6.2.6, page 33.
4
LC venting,
Interface
G1/2” for the venting hose
Connect the venting hose to this 1/2 ´´thread or to the equal threat on the top
side of the filling block. For further information, refer to chapter 6.2.6, page 33.
5
Mains connection
, 3-pin socket
To connect the mains plug.
Pay attention to the label on the socket to get the according device variation
connection order.
For further information in the technical data,
refer to chapter 9.4.3.1, page 55 .
6
Pump operation switch,
rocker switch
To switch on/ off the coolant pump.
For further information refer to chapter 4, page 20.
7
Earth ,
connection bolt M6
Connects the device to the protective earth.
Tab. 3
Overview of interfaces
– device rear side.